PPCH-G™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 18
Table 3.
MAIN RUN screen display fields
PRESSURE1UNITMhzRRH
TPRESSURE2 CC NN/NN
DISPLAY FIELD
NAME
PURPOSE
CONTENTS
SECTION
1. PRESSURE1 Measured
pressure
Displays pressure
measured by active
Q-RPT or utility
sensor
Numerical pressure
value and sign
1.2.2.1
1.2.2.2
3.3.10
2. UNIT
Unit
of
measure
Identifies unit of
measure in which
PRESSURE1 and
PRESSURE2 are
displayed
Pressure unit of
measure abbreviation
3.3.2
3. M
Measurement
mode
Identifies
measurement mode
of displayed pressure
<a>
: absolute
<g>
: gauge
3.3.3
4. h
Head
pressure
indicator
Indicates whether a
fluid head correction
is applied to
PRESSURE1
<h>
:
the fluid head
is not zero
<blank>
: fluid head is
zero
3.3.7
5. z
AutoZero
indicator
Indicates whether
the AutoZero
function is ON or
OFF
<z>
:
AutoZ is ON
<blank>
: AutoZ is OFF
3.5.1
6. RRH
Active
Q-
RPT position
indicator
Indicates the position
in the PPCH-G
system of the active
Q-RPT or utility
sensor
<IH>
: Internal
Hi
<IuH>
: Internal Hi
(utility
sensor)
<IL>
: Internal
Lo
<X1H>
: External 1 Hi
<X1L>
: External 1 Lo
3.2.5
7. NN/NN
Sequence
progress
indicator
Indicates progress of
an ATest sequence,
during test execution
<NN/NN>
: Number
of
this point over total
number of points in the
sequence
3.3.6
8. CC
Pressure
control
indicator
Indicates type of
pressure control,
whether control is
currently active and
whether custom
control limits are in use
<D>
: Standard dynamic
<d>
: Hi volume
dynamic
<S>
: Static
<M>
: Monotonic
<R>
: Rate
<P>
: Piston gauge
<C>
is appended to the
control character if
control limits are custom
Control character(s)
flash when PPCH-G is
actively controlling
3.2.1
3.4.6