PPCH-G™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 58
3.4.1.1
SAVING AN A
UTO
R
ANGE
RANGE
OPERATION
To save a range created by AutoRange, the range must be the active range.
Use AutoRange to create the range (see Section 3.3.4) and then make any
desired feature and setting adjustments.
Once the AutoRange range is set up as desired, press
[SETUP]
,
<1range>
and
select
<1save>
. The display is:
1.
Indication that this is an AutoRange range to
be saved.
2.
Designator and position of the Q-RPT used by
the range to be saved.
3.
Pressure unit of measure of the range to be
saved.
4.
Full scale pressure of the range to be saved in
gauge mode and absolute mode if available.
Save range A100M IH
MPa 80g/80a
Press
[ENT]
to save the range or
[ESC]
to return to the MAIN RUN screen without
saving the range.
Default (DF) Q-RPT ranges cannot be saved. They are permanent.
Duplicate AutoRange ranges cannot be saved. A duplicate AutoRange range is a
range that has the same measurement mode and full scale and uses the same Q-RPT
as a range that has already been saved.
3.4.1.2
DELETING A
UTO
R
ANGE
RANGES
OPERATION
To delete a single saved AutoRange range, press
[SETUP]
,
<1range>
,
<2delete>
. If there are any saved ranges available to delete, the display is:
1.
Indication that the cursor control keys may be
used to toggle through the saved ranges to
select the one to be deleted.
2.
Designator and position of the Q-RPT used by
the range to be deleted.
3.
Pressure unit of measure of the range to be
deleted.
4.
Full scale pressure of the range to be deleted
in gauge mode and absolute mode if available.
< and > A70M X1H
MPa 50g/50a
Use the cursor control keys to toggle through the ranges available to be deleted.
When the desired range to be deleted is displayed, press
[ENT]
. If you are sure you
want to delete this range, press
[ENT]
again in response to the
<Delete?>
query.
Press
[ESC]
to return to the
<AutoRange:>
screen without deleting a range.
To delete all of the saved AutoRange ranges, press
[SETUP]
,
<1range>
,
<3delete all>
. If you are sure you want to delete all the ranges, respond
<1yes>
to the
<Delete all saved ranges?>
query. If not, respond
<2no>
.
3.4.2
<2RES> (RESOLUTION)
PURPOSE
To set the resolution with which measured pressures and other pressure indications and
settings are displayed.