3. OPERATION
Page 51
© 2007 DH Instruments, a Fluke Company
Leak check is range specific in the sense that leak check is run using the active range.
However, only one set of leak check results is maintained in memory and each leak test completed
overwrites the memory. View leak check always shows the results of the last leak check run
regardless of the range that is now active. The results screen includes the range indicator to
indicate the range in which the leak check was run.
3.3.10
[ENT/SET P] (SET PRESSURE AUTOMATICALLY)
PURPOSE
To enter and execute target pressure and ramp values for PPCH-G automated pressure control.
See
Section
3.2.2 for information on automated pressure control and the different control
modes available with PPCH-G.
OPERATION
The operation of
[ENT/Set P]
depends on the active pressure control mode.
Dynamic mode:
See Section 3.3.10.1.
Static mode:
See Section 3.3.10.2.
Monotonic mode:
See Section 3.3.10.3.
Ramp mode:
See Section 3.3.10.4.
Interrupting automated pressure control:
See Section 3.3.10.5.
Commands for setting zero pressure:
See Section 3.3.10.6.
The active range pressure control mode and parameters can be viewed and edited using
[SPECIAL], <6control> (see Section 3.4.6).
3.3.10.1
DYNAMIC CONTROL MODE, [ENT/SET P]
If the PPCH-G is not set for dynamic control, use
SPECIAL, <6control>
,
<2mode>
to select dynamic control mode. Select
<1standard>
or
<2hi
volume>
depending upon the characteristics of the system into which PPCH-G
is controlling pressure (see Section 3.2.2.2).
For best results, wait 5 minutes with the PPCH-G vented before controlling
pressure when changing to or from standard dynamic control mode. This is to allow
time for the PPCH-G thermal pressure control unit to respond and stabilize.