FPG8601™/VLPC™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 4
1.3.2
VLPC GENERAL SPECIFICATIONS
Power Requirements:
85 to 264 VAC, 47 to 60 Hz, 70 VA max. consumption
Operating Temperature Range:
15 to 35 °C
Storage Temperature Range:
- 20 to 70 °C
Weight:
41.6 kg (91.5 lb)
Dimensions:
30.5 cm H x 53.4 cm W x 50.5 cm D
(12 in. H x 21 in. W x 19.9 in. D)
Microprocessor:
Motorola 68302, 16 MHz
Communication Ports:
RS232
IEEE-488.2
COM1: FPG
COM2:: Not Used
Not used
Control Specifications:
±
0.02 Pa + 30 ppm of range selected
Operating Medium:
Gas: air or Nitrogen (N
2
)
Pressure Supplies:
SUPPLY port: 700 - 850 kPa (100 - 120 psi), supply must remain
constant
±
1 % to meet control specifications
Drive Air Supply:
600 to 1000 kPa (90 to 150 psi) air
Pressure Connections:
Test (Lo) 1/4 VCO
Test (Hi) 1/4 VCO
ATM 1/4
VCO
Vacuum 3/8 VCO
Supply 1/8
NPT
F
Drive 1/8
NPT
F