FPG8601™/VLPC™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 94
FEATURE
DESCRIPTION
Model
(required drop
down list
selection)
The model number or name for the DUT can be selected from the drop
down list or a new model can be entered manually. The default Data File
directory naming preference can use this entry to create the data directory
for Date Files created when this DUT is run.
Serial Number
(optional text entry
field, Individual DUT
Definitions only)
Write in the DUT serial number. DUT Profiles do not include a serial
number during the creation of the DUT. The default Data File directory
naming preference can use this entry to create the data directory for Date
Files created when this DUT is run.
Identification
(optional text entry
field, Individual DUT
Definitions only)
Write a DUT identification, if desired. This field can be used for internal
tracking of DUTs or to record other information. Any combination of
information can be entered into the ID field. The value entered will be
included in the test Data Files. The default Data File directory naming
preference can use this entry to create the data directory for data files
created when this DUT is run. Profile Type DUT Definitions do not include a
device ID.
6.8.2.2
[CORRECTION] TAB
The
[Correction]
DUT Definition tab is provided to accommodate DUTs that can
use a linear correction applied to their output to achieve a higher accuracy. Enter
the adder, multiplier and calibration date in the fields provided. The adder and
multiplier are directly applied to the DUT output prior to calculating the DUT pressure.
As a result, the adder must be specified in the same units as the DUT output.
The slope and offset values should be determined by using a first order linear
regression (best fit). Enter a 1.0 for the <Slope> and 0 for the <Offset> to
disable this feature.
Figure 49.
DUT Definition Editor,
[Correction]
Tab
Table 36.
DUT Definition Editor,
[Correction]
Tab Fields
FEATURE
DESCRIPTION
Slope
(Optional numeric
entry)
Enter the multiplier to apply to the DUT output prior to calculating the DUT
pressure. Enter 1 to effectively disable this feature.
Offset
(optional numeric
entry)
Enter the offset to apply to the DUT output. The offset must be entered in
the output units.
Calibration Date
(optional text entry
field)
Calibration date associated with the slope and offset entries. This is a text
entry, so any date format can be accepted. The value entered is stored in
the data file for reference.