ADCS-601-AF™ OPERATION AND MAINTENANCE MANUAL
© 2005 - 2007 DH Instruments, a Fluke Company
Page 78
Figure 37
.
ADCS System Display
4.3.5.3
ADCS RUN SCREEN TOOLBAR
The
ADCS Run Screen Toolbar
is at the bottom of the
ADCS Run Screen
. Its
buttons are summarized inTable 17.
Table 17
.
ADCS Run Screen Toolbar
BUTTON
DESCRIPTION
Vent
Causes ADCS-601-AF to vent the
TEST(+)
port to atmosphere
(see Section 4.2.4.2).
Always makes PG6701 the active pressure standard and vents
unless the
<Use the FPG measurement when vented in gauge
mode>
option is active (see Section 4.2.4.1).
Abort Control
Aborts on-going pressure control. When the FPG8601 is the active
pressure standard, the pressure will continue to increase until a
new target pressure is set or the
[Vent]
button is pressed. The
increase in pressure is a direct result of the lubrication flow in the
FPG8601.
Always vent or set a new target pressure after aborting pressure
control when the FPG8601 is the active pressure standard.
Failure to do so will eventually result in an FPG8601 overpressure
condition. ADCS Tools automatically vents all system pressure
after an FPG8601 overpressure.
Change Pressure Unit of
Measure
Changes pressure unit of measure for all ADCS-601-AF pressure
values other than atmospheric pressure and reference vacuum.
Click any displayed quantity on any ADCS-601-AF run screen to
access a popup menu with a choice to change the unit of measure.
Change Active Pressure
Standard
Transition active pressure standard from PG7601 to FPG8601 or
vice-versa (see Section 4.1.3).
Since ADCS-601-AF automatically transitions pressure standards
as needed depending upon the target pressure request, this
function is not frequently used in normal operation.
Exercise
Setup and run an automated routine to cycle pressure a given
number of times between a mininimum and maximum value (see
Section 4.2.5).