When the test is complete the
Test Complete
form displays. If automatic report
generation is enabled (see Section 4.4.2.2), the
Report Editor
creates a report.
4.4.6.2
RUNNING A STANDARDIZATION
When the ADCS-601-AF is connected to the model 3682 ADC, do not execute any
function on the 3682 that will cause it to vent pressure unless the ADCS-601-AF
is vented first. Causing pressure on the ADCS-601-AF TEST(+) port to vent when
the ADCS-601-AF is not vented, may cause damage to the ADCS-601-AF.
NOTE: The ADCS-601-AF is always vented before the model 3682 vents if the
procedures below are followed as instructed.
The model 3682 standardization procedure steps the operator through the model
3682 required sequence. This procedure cannot be fully automated due to
limitations in the model 3682 remote validation procedure.