ADCS-601-AF is a fully automated reference level Air Data Calibration Standard for the verification and
calibration of air data range instruments, in particular Air Data Test Sets (ADTS). ADCS-601-AF’s intended
function is the static pressure calibration of Pt and Ps range pressure transducers with very low measurement
uncertainty. ADCS-601-AF is not intended to duplicate the functions of an ADTS in calibrating and testing on-
board aircraft instruments.
ADCS-601-AF covers the pressure range of 1 to 380 kPa (0.3 to 120 inHg) in absolute and gauge modes
with fully automated operation and state of the art uncertainty.
ADCS-601-AF is an integrated system made up of several instruments that work together under the control of
ADCS Tools software running on a personal computer. Operator interaction with ADCS-601-AF is
accomplished through the keyboard, pointing device and display of the system controller (personal computer).
The heart of the system is a fully automated PG7601 piston gauge or pressure balance that covers the
range of 15 to 380 kPa (4.5 to 120 inHg). The PG7601 range is extended down to 1 kPa (0.3 inHg) by an
FPG8601 force balanced piston gauge with automated pressure control.
1.1
SPECIFICATIONS
1.1.1
ADCS SYSTEM GENERAL SPECIFICATIONS
Power Requirements
Instruments (AC1)
Reference vacuum pumps (AC2)
Utility pump (AC3)
100 to 240 VAC, 50/60 Hz, 240 W max. consumption
Fuse: 250VT 10A
110 V version: 100 – 120V, 50/60 Hz, 1100 W max. consumption
Fuse: 250VT 10A
220 V version: 200 – 240V, 50/60 Hz, 1100 W max. consumption
Fuse: 250VT 10A
110 V version: 100 – 120V, 50/60 Hz, 800 W max. consumption
Fuse: 250VT 10A
220 V version: 200 – 240V, 50/60 Hz, 800 W max. consumption
Fuse: 250VT 10A
Pressure Supply Requirements
Drive air: 0.7 to 1 MPa (100 to 150 psi)
Test gas: Nitrogen or air, 0.45 to 1 MPa (65 to 150 psi), clean and dry
(instrument grade minimum, high purity preferred)
Operating Temperature Range
20 to 26 ºC
Weight
Reference bench
Control cabinet
Utility and roughing vacuum pumps
173 kg (380 lb) approx. (with turbo pump and controller installed)
105 kg (230 lb) approx.
19 kg (42 lb) each
Dimensions
Reference bench
Control cabinet
145 cm H x 122 cm W x 76 cm D (57 in. x 48 in. x 30 in.)
124 cm H x 56 cm W x 66 cm D (49 in. x 22 in. x 26 in.)
Pressure Ranges
Overall
PG7601
FPG8601
1 to 380 kPa (0.3 to 120 inHg)
10 to 380 kPa (3 to 120 inHg)
1 to 15 kPa (0.2 to 4.5 inHg)
Operating Medium
Nitrogen or air
Pressure Connections
TEST(+) port
TEST(-) port
Quick connector (1/4 in. NUPRO
®
SS-QC4-B1-400)
Quick connector (1/4 in. NUPRO
®
SS-QC4-B1-400)
CE Conformance
Available, must be specified.