ADCS-601-AF™ OPERATION AND MAINTENANCE MANUAL
© 2005 - 2007 DH Instruments, a Fluke Company
Page 90
The
Data Grid Run Screen
is available any time a test is running for which there
is a Data File. The screen vizualizes the data taken during the test, point by
point in columns.
The
Data Grid Run Screen
can be opened by pressing the
[Data Grid]
button
on the
ADCS Tools Main Tool Bar
(see Section 4.3.3, Table 14).
The contents of the
Data Grid Run Screen
can be customized using the
[Data
Grid]
tab of
[Tools]
,
[Test Options]
(see Section 4.4.1.4).
Figure 43.
Data Grid Run Screen
4.4.3.3
DUT/REFERENCE COMPARISON RUN SCREEN
ADCS Tools uses the engine of COMPASS for Pressure® calibration software to
manage automated testing using ADCS-601-AF. The DUT/Reference
Comparison Run Screen supported by ADCS Tools is a COMPASS for Pressure
feature. See the COMPASS for Pressure User’s Manual for additional
information on this screen and its options. The COMPASS for Pressure manual
is an on-line manual available by pressing the [?] button, when available, in
COMPASS features. It can also be accessed on the ADCS-601-AF system
controller in the COMPASS for Pressure program group reached using the
Windows START button , Programs.
The
Reference/DUT Comparison Run Screen
is available any time a test is
running. The screen vizualizes real time DUT and ADCS-601-AF readings and
the error of the DUT(s) compared to the ADCS-601-AF. The information on the
display is organized into rows and columns. The columns separate different
pieces of information. The rows separate the different devices. The first row is a
label that describes the output. The second row is always the ADCS-601-AF
output information. All subsequent rows are for the number of DUTs in use..
The
Reference/DUT Comparison Run Screen
can be opened by pressing the
[Reference/DUT Comparison]
button on the
ADCS Tools Main Tool Bar
(see
Section 4.3.3, Table 14).
Figure 44.