4. OPERATION
Page 95
© 2005 - 2007 DH Instruments, a Fluke Company
Press
[OK]
only when the model 3682 is in
<Calibration Manual >
mode
and the
<PS>
pressure is updating on its front panel display.
ADCS Tools
provides a prompt of whether to exercise the transducer.
Select
[YES]
to exercise the transducer by applying full scale pressure three
times with 10 second dwell. Select
[NO]
to skip the exercise step.
The Offset Check Point pressure of 28.8 inHg is set.
When Ready occurs at the Offset Check Point pressure a popup displays
indicating the actual pressure set by ADCS-601-AF.
On the front panel of the model 3682 perform the following operations:
•
Press
[MEASURE]
. The
[ACTUAL]
field begins to blink
•
Enter
the
<ADCS-601-AF Pressure>
displayed on the
ADCS Tools
popup using the numeric keys at the bottom of the model 3682 display.
Verify that the information is entered correctly and press
[Enter]
.
•
The model 3682 prints the calibration offset value. Enter this value
in the edit field at the bottom of
ADCS Tools
popup. The value is
stored in the test data file and printed out on reports.
•
Press
[OK]
to continue. The ADCS-601-AF vents the pressure
immediately after
[OK]
is pressed. It is important that the
ADCS-601-AF vent rather than the model 3682 venting the pressure.