13
A-306
EN
cAring for your igo system
SYMPTOM
POSSIBLE.CAUSE
REMEDY
Unit does not operate when using
with charged battery. Power Fail
alarm activates.
1. Battery is overheating.
1. Allow battery to cool.
DC Adapter is attached. Unit runs
from Battery or does not run. The
External Power Present Light not
illuminated.
1. Poor connection to DC power source.
2. Vehicle power source (vehicle accessory
connector) dropped too low for the DC
Adapter.
3. Fuse in vehicle is blown.
1. Ensure the DC accessory outlet is clean and a good
connection can be made. Insert the other end of the DC
Adapter into the vehicle's DC accessory outlet.
2. If the vehicle power source drops too low for the DC
Adapter, the iGo will revert to Battery operation (if
present) or will cease operation until power is restored.
3. Check the fuse and replace if necessary.
Lowest Battery Status Light is
Yellow. Unit beeps once.
1. Battery needs to be charged.
1. Plug unit into AC power or replace discharged battery
with charged battery.
2. Plug unit into DC accessory outlet to operate unit
(battery will not recharge on DC power).
Battery Status Lights never
indicate fully charged.
1. Battery needs to be conditioned.
2. Defective battery.
1. Fully discharge battery then recharge.
2. Contact your DeVilbiss provider.
Yellow Battery Status light is
flashing.
1. Unit is equipped with built-in battery test
and the battery pack is being checked.
1. If light flashes longer than 5 minutes, battery is defective.
Contact your DeVilbiss provider for a replacement.
Audible Alert sounds
intermittently when operating
from DC Adapter.
1. Vehicle not running.
2. Poor connection to DC power source.
3. Vehicle electrical system overloaded or
defective.
1. Start vehicle.
2. Ensure the DC accessory power port outlet is clean and
a good connection can be made. Insert the DC Adapter
into the vehicle’s DC accessory outlet.
3. Have a qualified auto technician check electrical system
while iGo is attached.
Short Battery Run Time.
1. Battery needs to be reconditioned.
2. Air filter is blocked.
3. Exhaust is blocked.
4. Intake filter silencer is blocked.
1. Perform a complete discharge and then a 100% charge
of the battery.
2. Check the air filter. If the filter is dirty, wash it following
the cleaning instructions.
3. Check the exhaust area; make sure there is nothing
restricting the unit exhaust.
4. Replace the intake silencer/filter.
If the above remedies do not work, contact your DeVilbiss
provider.
If any other problems occur with
your iGo.
1. Turn your unit Off and switch to your reserve oxygen
system. Contact your DeVilbiss provider immediately.
CaRInG foR YoUR DevIlbIss IGo PoRTable oXYGen sYsTeM
�
WArning
Before.attempting.any.cleaning.procedures,.turn.the.unit.Off..
NOTE
-Use no lubricants, oils, or grease.
Air.Filter
The air filter should be cleaned at least once a week. To clean, follow these steps:
1. Remove the air filter.
2. Wash in a solution of warm water and dishwashing detergent.
3. Rinse thoroughly with warm tap water and towel dry. The filter should be completely dry before reinstalling.
.
CAUTION
-To prevent product damage, do not attempt to operate the unit without the air filter or while the filter is still damp.
AC.Adapter.and.Power.Cord/DC.Adapter
The adapters and power cords should be cleaned as needed. To clean, follow these steps:
1. Disconnect cords from the iGo before cleaning.
2. Clean the cords by using a damp cloth with a mild household cleaner and wipe dry.
Cannula.and.Tubing
Clean and replace the cannula and tubing as instructed by your DeVilbiss provider.