
50
1.2 Technical data
Note:
The units Multimat2 Touch und Multimat2 Touch+Press are delivered in 2 different voltage systems,
including a power lead (as required for the country in which it is operated):
• 100 – 125 Volt ~ / 50/60 Hz
• 230 – 240 Volt ~ / 50/60 Hz
Check the voltage rating of your unit on the type plate, which is located on the rear of the unit. Make sure that the
voltage rating and the supplied power cord are appropriate for your local power network.
Multimat2 Touch
Multimat2 Touch+Press
Power consumption
1,350 W (without vacuum pump)
1,350 W (without vacuum pump)
Pump connection
As mains voltage, max. 2.5 A
As mains voltage, max. 2.5 A
Overvoltage
category
II
II
Protection
class
I
I
Instrument fuses
100 — 125 V : 2 x 10 AT
100 — 125 V : 2 x 10 AT
230 — 250 V : 2 x 6.3 AT
230 — 250 V : 2 x 6.3 AT
Degree
of
contamination
2
2
Weight
22 kg
25 kg
Dimensions H/W/D
441/320/425 mm closed
593/320/425 mm closed
Dimensions H/W/D
585/320/425 mm opened
748/320/425 mm opened
Input compressed air pressure
–
2,7 bar
1.3 Environment conditions
Environment temperature:
2 °C to 40 °C (35,6 °F to 104 °F)
Relative humidity:
max. 80 % at 31 °C (87,8 °F)
Altitude:
max. 3500 m above sea level
Summary of Contents for Multimat 2 Touch
Page 2: ......
Page 42: ...42 ...
Page 82: ...82 ...
Page 122: ...122 ...
Page 162: ...162 ...
Page 203: ...203 ...
Page 204: ...www dentsply degudent com LU c ...