• Incoming water temperature to the Cavitron System
should not exceed 25˚C (77˚F). If needed a device
should be installed to maintain a temperature within this
specification, or a Cavitron DualSelect Dispensing
System attached to allow this system to be operated
as a closed water system.
7 2 Electrical Requirements
• Incoming power to the system must be 100 volts AC to
240 volts AC, single phase 50/60 Hz capable of
supplying 1.0 amps.
• The system power should be supplied through the AC
power cord provided with your system.
• WARNING: To avoid risk of electric shock, this equipment
must only be connected to a supply mains with protective
earth.
7 3 Unpacking the System
Carefully unpack your Cavitron Plus Ultrasonic Scaler and
verify that all components and accessories are included:
1. Cavitron
®
Plus
™
System with Handpiece Cable
Assembly with swivel
2. Water Line Assembly (Blue) with Filter and Quick
Disconnect
3. Additional Water Line Filter
4. Detachable AC Power Cord
5.
Cavitron
®
Tap-On
TM
Wireless Foot Pedal
6. “AA” Batteries (4-Pack)
7. Auxiliary Cable for Tap-On
TM
Foot Pedal
8. Steri-Mate
®
Grip (not shown)
9. Steri-Mate
®
Detachable Sterilizable Handpiece
10. Cavitron
®
Ultrasonic Inserts (quantity optional)
11. Efficiency Indicator for Cavitron Inserts
12. Literature Packet
7 4 System Installation
• The Cavitron Plus System is designed to rest on a level
surface. Be sure unit is stable and resting on four feet.
• The Cavitron Plus System should not be positioned such
that access to the power input and AC power cord are
limited.
• Placing unit in direct sunlight may discolor plastic
housing.
• The system has been equipped with a Tap-On
TM
Wireless
Foot Pedal which was factory synchronized to operate
with the system’s base unit. If your office has more
than one Cavitron system with Tap-On
TM
Technology,
it is recommended that you mark the Tap-On
TM
Foot
Pedal and base unit for easy reference as to which Tap-
On
TM
Foot Pedal operates with which base unit. Should
resynchronization be necessary, follow the instructions in
section 7.10.
7 5 Power Cord Connection
• Verify the Main Power ON/OFF switch, located at the
center front underside of the System, is set to the OFF
(O) position before proceeding.
• Insert the AC power cord into the power input on the
back of the System.
• Insert the pronged plug into an AC wall outlet.
7 6 Water Supply Line Connection
• Grasp the Water Supply Line (blue hose) by the end
opposite the quick-disconnect and insert it into the water
inlet connector until fully seated.
Press ring to release
water supply tube.
Summary of Contents for Cavitron Plus Ultrasonic Scaler
Page 1: ......
Page 3: ......
Page 4: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 19: ......
Page 25: ......
Page 26: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 43: ......
Page 49: ......
Page 50: ......
Page 58: ......
Page 59: ......
Page 60: ......
Page 61: ......
Page 62: ......
Page 66: ......
Page 73: ......
Page 74: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 90: ......
Page 96: ......
Page 97: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 113: ......
Page 120: ......
Page 121: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 138: ......
Page 143: ...143...
Page 144: ......