23
ACTION TAKEN
1.
Check that the Main Power Switch is in the ON (I) position, and that the detachable power cord is fully seated in the
receptacle on back of system.
2.
Check that the system’s power cord plug is fully seated in an appropriate AC wall outlet.
3.
Check that the wall outlet is functional.
1.
If the office has more than one Tap-On
TM
Foot Pedal, test each to ensure that the proper Tap-On
TM
Foot Pedal is being
used. With a handpiece and insert installed, depress the Tap-On
TM
Foot Pedal to the first position. The system should
dispense water. If none of the Tap-On
TM
Foot Pedals operate the system, continue to the next step.
2.
Resynchronize one Tap-On
TM
Foot Pedal to the system (see Section 7.10 Tap-On
TM
Foot Pedal Synchronization).
1.
Assure that handpiece lavage control is properly adjusted.
2.
Check for clogged insert. Replace insert if necessary.
3.
Check that dental office water supply valves are open.
4.
If the system is connected to DualSelect Dispensing System, check that fluid level in the selected bottle is sufficient.
Make sure valves are open when using external water source.
5.
Check that the water line filter is clean. Replace filter, if needed.
1.
Check that the Power Control is not in Rinse Mode.
2.
Check the insert for damage and that it is properly installed in the Handpiece.
3.
Check that the handpiece is properly installed to the cable assembly.
4.
Verify that the soft nozzle grip is flush with the hard plastic of the insert port.
5. Turn the system’s Main Power Switch OFF, (O) position. Wait 5 seconds and turn the system back ON.
6.
If problem still exists, replace both “AA” batteries in Tap-On
TM
Foot Pedal with new “AA” batteries (Refer to Section 7.9) or
connect Auxiliary Tap-On
TM
Foot Pedal Cable.
1.
Fast blinking (3 blinks per second) – Indicates improper set-up
A. If insert is in the handpiece, remove. Verify the handpiece is properly seated and hold the Tap-On
TM
Foot Pedal for
2 seconds. If blinking stops, the system is ready for use. If blinking remains, continue to the next step.
B. Attach a NEW handpiece and depress Tap-On
TM
Foot Pedal for 2 seconds. If blinking stops, the system is ready
for use. Discard the old handpiece or return if within warranty. If blinking remains, continue to the next step.
C. Install and fully seat an insert into handpiece. Depress Tap-On
TM
Foot Pedal for 2 seconds. If blinking stops, the
system is ready for operation. If blinking remains, continue to the next step.
D. Install and fully seat a NEW insert in handpiece and depress Tap-On
TM
Foot Pedal for 2 seconds. If blinking stops,
system is ready for use. Discard old insert or return if within warranty. If blinking remains, refer to Section 11.2
Technical Support and Repairs to have unit serviced as soon as possible.
2.
Slow blinking (1 blink per second) - The system is not operating within factory specifications.
A.
Remove insert.
B.
Turn Main Power Switch OFF, (O) position. Wait five seconds. Turn unit ON, (I) position.
C.
Operate Purge function.
D.
If service indicator still blinks, refer to Section 11.2 Technical Support and Repairs to have unit serviced as soon
as
possible.
1.
Ensure that the base unit has adequate ventilation and is not near a heat source (i.e. radiator, heat lamp, sunlight or other
heat producing operatory equipment).
2.
Turn Main Power Switch OFF (O) position. Allow system to cool for 10 minutes and turn system to the ON (I) position.
Verify light is not illuminated.
3.
If light is still illuminated, refer to Section 11.2 Technical Support and Repairs to have unit serviced as soon as possible.
1.
Check that there is no insert in the handpiece.
2.
Check that Handpiece is properly installed to the cable assembly.
QUICK REFERENCE GUIDE: TROUBLESHOOTING
SYMPTOM
System will not
operate: No Power
ON Indicator
System will not
operate: Power ON
Indicator is illuminated
System operates:
No water flow
to insert tip or
handpiece
overheats
System operates:
No insert
cavitation
System operates:
Service
indicator
blinking
System operates:
Service indicator
illuminated
System operates:
Purge mode will not
function -
icon flashing
Summary of Contents for Cavitron Plus Ultrasonic Scaler
Page 1: ......
Page 3: ......
Page 4: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 19: ......
Page 25: ......
Page 26: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 43: ......
Page 49: ......
Page 50: ......
Page 58: ......
Page 59: ......
Page 60: ......
Page 61: ......
Page 62: ......
Page 66: ......
Page 73: ......
Page 74: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 90: ......
Page 96: ......
Page 97: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 113: ......
Page 120: ......
Page 121: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 138: ......
Page 143: ...143...
Page 144: ......