22
Recommended separation distance between Portable and mobile RF communications
equipment and the model @ 3Vrms
The model G136 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Model G136 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Model G136 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
- 0.12
0.23
0.1
- 0.34
0.74
1
- 1.7
2.3
10
- 3.7
7.4
100
- 11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Recommended separation distance between Portable and mobile RF communications
equipment and the model @ 10Vrms
The model G136 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Model G136 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Model G136 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
- 0.035
0.07
0.1
- 0.11
0.22
1
- 0.35
0.7
10
- 1.1
2.2
100
- 3.5
7
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Summary of Contents for Cavitron Plus Ultrasonic Scaler
Page 1: ......
Page 3: ......
Page 4: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 19: ......
Page 25: ......
Page 26: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 43: ......
Page 49: ......
Page 50: ......
Page 58: ......
Page 59: ......
Page 60: ......
Page 61: ......
Page 62: ......
Page 66: ......
Page 73: ......
Page 74: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 90: ......
Page 96: ......
Page 97: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 113: ......
Page 120: ......
Page 121: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 138: ......
Page 143: ...143...
Page 144: ......