Trigno
TM
Personal Monitor User’s Guide
37
1) Data Streaming
This is the typical data transmission state, where a sensor is commu-
nicating with the TPM receiver and all data packets are being prop-
erly received. This state is indicated by a constant green blink every
second.
2) Receiver Scanning
This state indicates that the sensor is not transmitting data to the re-
ceiving unit, and more specifically not receiving any information from
the base. This can happen if the senor is out of range or if the base sta-
tion has become inactive (powered off or entered a sleep mode), and
is denoted by an alternating green then amber flash once per second.
3) Sensor Pair Complete
Three successive green LED flashes indicate that the pairing opera-
tion between the sensor and the receiving unit has successfully been
completed.
4) Sensor Pair Incomplete
Three successive red LED flashes indicate that the initiated pair com-
mand did not complete correctly, most likely because the sensor is too
far from the base station.
5) Battery Charging
When docked in the recharge station, sensors will show an amber
LED during recharging..
6) Charge Complete
When docked in the recharge station, sensors will show a green LED
when their charge cycle is complete.
7) Charge Error
When docked in the recharge station, an LED that is off indicates a
charge error. Re-seat the sensor to ensure a proper connection with
the charge contacts and to restart the charge cycle. Contact Delsys
technical support if the problem persists after re-seating the sensor.
8) Sensor Off
When out of the recharge station, a sensor with no LED illumination
is powered off.