Trigno
TM
Personal Monitor User’s Guide
31
Resetting the Device
Initialization
If the battery is fully depleted, initial configuration is needed to re-
store device settings. To perform this configuration, plug the USB
cable into any PC with EMGworks installed and launch EMGworks.
When EMGworks is closed, the device will be fully configured.
Figure 31:
Initializing the TPM with EMGworks.
System Clock Reset
The TPM will automatically synchronize its clock to the system clock
of any PC to which it is connected whenever EMGworks is launched.
The clocks can also be manually synchronized via a button click in
EMGworks.
Hard Reset
Note: A hard-reset of the device may cause internal memory cor-
ruption. Consultation with Delsys technical support prior to initi-
ating a hard reset is recommended.
If the device becomes non-responsive, it may be powered off with a
hardware reset sequence. Depending on the state of the TPM when
the hard-reset is initiated, the hardware reset may cause unintended
corruption to the device’s persistent memory. Use care when initiat-
ing a hard-reset.
The hard reset is performed by pressing softkey 1, softkey 3, and se-
lect/enter simultaneously. When these three buttons are depressed,
power is immediately cut to all portions of the device.
After the hard reset, the device may be powered on normally.