venTest OPERATORS MANUAL
Range and Accuracy
Parameter
Units
Range
Accuracy
(whichever is
greater)
Plateau
2
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 1% or 0.02 kPa* **
2
Mean
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 1% or 0.02 kPa* **
I/E
300/1 - 1/300
± 0.1 **
Ti
s; ms
0 - 60 sec
± 1% ± 0.02sec **
Te
s; ms
0 - 60 sec
± 1% ± 0.02sec **
Tip
s; ms
0 - 60 sec
± 1% ± 0.02sec **
Tep
s; ms
0 - 60 sec
± 1% ± 0.02sec **
Humidity
% rh
5 - 95%
5 - 80 % rh: ± 3%
80 - 95 % rh: ± 4%
Insp. Peak Flow
l/min; ml/min
0 - 300 L/min (AIR)
2
0 - 200 L/min (O ,
2
AIR/O
± 2% of m.v. or 0.1
l/min (AIR) **
± 4% of m.v. or 0.2
2
2
l/min (O , AIR/O ) **
Exp. Peak Flow
l/min; ml/min
0 - 300 L/min (AIR)
2
0 - 200 L/min (O ,
2
AIR/O
± 2% of m.v. or 0.1
l/min (AIR) **
± 4% of m.v. or 0.2
2
2
l/min (O , AIR/O ) **
2
HFO Pmean
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 2% of m.v. or 0.04
kPa* **
2
HFO PMax
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 2% of m.v. or 0.04
kPa* **
2
HFO PMin
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 2% of m.v. or 0.04
kPa* **
2
HFO amplitude
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 2% of m.v. or 0.04
kPa* **
HFO frequency
Hz
0 - 20
± 1 Hz
HFO I/E
3/1 - 1/3
± 0.3
(*) Please note that due to the pneumatic resistance of the Flowsense 300 Sensor the measured pressure from the ventilation pressure sensor may differ from values
measured at other places in breathing system, especially at high flows .
(**) Subject to trigger settings
Specifications/Chapter 1
#
Page 3