venTest OPERATORS MANUAL
1.1.2 FlowSense 012 (Low Flow) Sensor Measurement
Physical Data
Dimensions
125 mm x 41 mm x 58 mm (L x W x H)
Connector 1
6 mm tube connection
Connector 2
6 mm tube connection
Weight
Approx. 290 g
Environmental Conditions
Parameter
Operation
Storage
Temperature
10 to 40
E
C
-20 to 50
E
C
Ambient Humidity
5 to 95% RH
(non-condensing)
5 to 95% RH (non-condensing)
Gas Temperature
10 to 40
E
C
Gas Humidity
0 to 95% RH
(non-condensing)
Ambient Pressure
60 - 110 kPa
60 - 110 kPa
Range and Accuracy
Parameter
Units
Range
Accuracy
(whichever is
greater)
Flow: Air
± 12 l/min
± 2% or 0.01 l/min
(STPD)
2
Flow: O
± 12 l/min
± 4% or 0.02 l/min
(STPD)
2
Flow: Air/O
± 12 l/min
± 4% or 0.02 l/min
(STPD)
Vti
l; ml
0 - 7.5 L
± 2% of m.v. or 20 ml
(AIR)
‡
Vte
l; ml
0 - 7.5 L
± 2% of m.v. or 20 ml
(AIR)
‡
Specifications/Chapter 1
#
Page 4
Summary of Contents for venTest
Page 1: ...Operating Manual venTest Ventilator Tester MN 108 ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 32: ...venTest OPERATORS MANUAL Overview Chapter 2 Page 18 ...
Page 116: ...venTest OPERATORS MANUAL Appendix C Page 102 ...
Page 117: ......