venTest OPERATORS MANUAL
Range and Accuracy
Parameter
Units
Range
Accuracy
(whichever is
greater)
MVi
l/min; ml/min
0.5 - 12 L/min (AIR)
2
0.5 - 12 L/min (O ,
2
AIR/O )
± 2% of m.v. or 20
ml/min (AIR)
‡
± 4% of m.v. or 40
2
2
ml/min (O , AIR/O )
‡
MVe
l/min; ml/min
0.5 - 12 L/min (AIR)
2
0.5 - 12 L/min (O ,
2
AIR/O )
± 2% of m.v. or 0.2
l/min (AIR)
‡
± 4% of m.v. or 0.4
2
2
l/min (O , AIR/O )
‡
Breath Rate (Freq)
BrPM
1 - 100 BrPM
101 - 200 BrPM
± 0.1 BrPM
‡
± 0.2 BrPM
‡
Ventilation Pressure
kpa; Mpa; bar; mbar;
2
cmH O; mmHg; psi;
2
hpa; Torr; inH O; inHg
0 - 10 kPa
± 1% of m.v. or
0.2 hPa
2
PEEP
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 1% or 0.02 kPa
† ‡
2
Peak
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 1% or 0.02 kPa
† ‡
2
Plateau
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 1% or 0.02 kPa
† ‡
2
Mean
kpa; mbar; cmH O;
mmHg; psi; hpa
0 - 10 kPa
± 1% or 0.02 kPa
† ‡
I/E
300/1 - 1/300
± 0.1
‡
Ti
s; ms
0 - 60 sec
± 1% ± 0.02sec
‡
Te
s; ms
0 - 60 sec
± 1% ± 0.02sec
‡
Tip
s; ms
0 - 60 sec
± 1% ± 0.02sec
‡
Tep
s; ms
0 - 60 sec
± 1% ± 0.02sec
‡
Humidity
% rh
5 - 95%
5 - 80 % rh: ± 3%
80 - 95 % rh: ± 4%
Specifications/Chapter 1
#