venTest OPERATORS MANUAL
1.2
Volume Measuring with Pressure Triggering
Volume measurement with 2 pressure
trigger thresholds, P1 and P2.
Volume measurement is going to start with
pressure exceeding P1, ending with
pressure being below P2.
In volume calculation an existing flow-base
will be added.
1.3
Volume Measuring with Pressure/Flow Triggering
Volume measurement with a flow trigger
threshold F1 and a pressure trigger
threshold P2.
Volume measurement is going to start with
flow exceeding F1, ending with flow being
below P2.
In volume calculation an existing flow-base
will be added.