venTest OPERATORS MANUAL
1.1.4 IRMA AX+ Multigas Sensor
TM
†
Physical Data
Dimensions
38 mm x 37 mm x 34 mm (L x W x H)
Weight
Approx. 25 g
Dead Space: Adult
Infant
6ml
1ml
Warm-up Time
10 sec, full specification within 60 sec
Interface
RS-232 at 9600 bps
Environmental Conditions
Parameter
Operation
Storage
Temperature
5 to 40
E
C
-20 to 50
E
C
Ambient Humidity
10 to 95% RH
(non-condensing)
0 to 100% RH (non-condensing)
Gas Temperature
10 to 40
E
C
Gas Humidity
0 to 95% RH
(non-condensing)
Ambient Pressure
53 - 120 kPa
Range and Accuracy
Parameter
Units
Range
Accuracy
2
CO
Vol %
0 - 10
10 - 15
± 0.2vol% + 2% of
reading
± 0.3vol% + 2% of
reading
2
N O
Vol %
0 - 100
± 2vol% + 2% of
reading
HAL, ISO, ENF
Vol %
0 - 8
± 0.15vol% + 5% of
reading
SEV
Vol %
0 - 10
± 0.15vol% + 5% of
reading
DES
Vol %
0 - 22
± 0.15vol% + 5% of
reading
For further information refer to the Masimo Sweden Website (www.masimosweden.com)
†