Performance
WARNING: Operation of the monitor
outside the specified values may
cause inaccurate results.
Datex-Ohmeda S/5 FM
Power supply
Rated voltages and
frequencies:
100 to 240V 50/60 Hz
Allowed voltage fluctuations: ±10%
Max. power consumption: 150 VA
Battery operation
Batteries:
Exchangeable lithium-ion,
2 pcs
Charging time:
3 hours per battery pack
Operation time:
Up to 5 hours
Environmental conditions
Operating temperature:
normal operation: +10 to +40°C (50 to 104°F)
with CO
2
measu10 to +40°C
(50 to 104°F)
while charging batteries: +10 to +35°C
(50 to 95°F)
Storage and transport temperature:
-20 to +60°C (-4 to 140°F)
Relative humidity: 10 to 90% noncondensing,
in airway 0 to 100%
condensing
Atmospheric pressure: 660 to 1060 mbar
(500 to 800 mmHg)
Alarm behavior
The maximum alarm delay of the alarm at the
monitor signal output to network: 1.1 seconds
If the alarm mode is latched, the technical alarms
are latched as well. This does not comply with the
NIBP (IEC 60601-2-30) and invasive pressure
(IEC 60601-2-34) standard requirements.
Silencing alarms for 5 minutes does not comply with
the SpO
2
(ISO 9919) standard requirements.
Wireless Network Option, N-FMW
Type:
Built-in transceiver and
antenna
Frequency range:
Worldwide product covering 2.4 to 2.5 GHz,
programmable for different country regulations
Data rate:
11 Mbps per channel (max.)
Output power:
100 mW
Data transmission: IEEE 802.11b compliant, Direct
Sequence Spread Spectrum
(DSSS)
Security
Wired Equivalent Privacy (WEP)
40 and 128 bit encryption
Certificates: Wi-Fi-certified
N-FREC, N-FCREC
Power consumption: 3 W
Recorder type:
Thermal array
Print resolution:
Vertical 8 dots/mm (200 dots/inch)
Horizontal 32 dots/mm (800 dots/inch) at a
speed of 25 mm/s and slower
Paper width:
50 mm, printing width 48 mm
Traces: Selectable
1, 2, or 3 traces
Print speed:
1, 6.25, 12.5, 25 mm/s
Hemodynamic modules E-PSM,
E-PSMP
ECG
1)
Filter modes:
monitoring filter 0.5 to 30 Hz
ST filter
0.05 to 30Hz
diagnostic filter 0.05 to 150 Hz
With 60 Hz power supply frequency:
monitoring filter 0.5 to 40 Hz
ST filter
0.05 to 40 Hz
diagnostic filter 0.05 to 150 Hz
QRS minimum detection level:
Minimum level 0.5 mV with duration between
40 and 120 ms fulfils ANSI/AAMI EC13 standard.
Defibrillation protection: 5000 V, 360 J
Recovery time:
5 seconds
1)
The isolation barrier capacitance in the module
has been minimized to reduce the hazard of burns in
the event of a defect in the ESU return electrode
connection
.