Titration table
The titration table gives 50 titration values for the selected drug. The
monitor displays both infusion rates and drip rates. If required, both
the starting dose and the increment steps can be edited through the
Installation
menu; for more information, see the “User’s Reference
Manual.”
To have a titration table for a given drug:
1. Press the
Pt.Data & Trends
key.
2. Select
Drug Calculator
.
3. Select
Table
-
Edit Values
.
4. Enter the data you have available, for example the drug name and
concentration, by turning and pushing the ComWheel.
5. The monitor calculates the infusion rate and the drip rate.
Printing or recording the titration table
In some cases, the titration table may be required near the patient.
You can print the table on the recorder in modules N-FREC and
N-FCREC, or on a laser printer. The
Calculator
and
Table
values are
both printed at the same time.
1. Press the
Pt.Data & Trends
key.
2. Select
Drug Calculator
.
•
Select
Record Calcs
to print them on the recorder.
•
Select
Print Calcs
to print them on a laser printer.