Adjusting limits
1. Press the
Alarms Setup
side panel
key and select
Adjust Limits
.
2. Highlight the measurement.
3. Push the ComWheel to open an adjustment window.
4. Turn the ComWheel to change limits and accept them by pushing
it. Move between selections by turning the ComWheel.
NOTE: If the monitor is connected to the network, the alarm limits can
also be changed using the S/5 iCentral if this feature has been
enabled in the iCentral configuration.
Adjusting volume
1. Press the
Alarms Setup
side panel key.
2. Select
Alarm Volume
and adjust
.
NOTE: If the monitor is connected to the network, the alarms can be
heard and seen on the S/5 iCentral as well. Please, consult the "S/5
iCentral User’s Reference Manual:
Alarms
" for details.
Alarm light
The S/5 FM has an alarm light, located in the upper right corner of
the monitor frame, see page 9. The alarm light can be enabled
(default) or disabled through
Monitor Setup
-
Install/Service
-
Installation
–
Alarm Options
-
Alarm Light
. When enabled, it flashes
red or yellow according to the currently active highest priority alarm.
To adjust the brightness of the light:
1. Press the
Alarms Setup
key.
2. Select
Alarm Light
and adjust with the ComWheel. During
adjustment the light is on to help you determine a suitable
brightness level.
Changing source
For NIBP, P1 and P2 (with E-PSMP) and Temp, you can select which
measured values trigger the alarm. One or several alarm sources may
be active at a time.
1. Press the
Alarms Setup
side panel
key and select
Adjust Limits
.
2. Select the measurement.
3. If the highlight is in the adjustment window, push the ComWheel
until you get to the menu selections.
4. Select the desired alarm source ON
.
Receiving alarms from other sites
NOTE: The monitor needs to be connected to a network.
1. Press the
Pt.Data & Trends
key.
2. Select
Other Patients
.
3. Select
Receive Alarms
.
4. Select the correct network and one of the other sites.
Showing alarm history
•
Press the
Pt.Data & Trends
key.
•
Select
Alarm History
. A list of the last 20 alarms is displayed.