68
68
Starting
The monitor sets inflation limits automatically for adults and infants
according to the hose used. For children and when using hoses without
identification, the inflation limit must be set manually. To do this:
1.
Press the
NIBP
key and select
NIBP Setup
-
Inflation Limits
.
NOTE: When using hoses without identification, the monitor goes
to this selection automatically when you press the
NIBP
key. With
these hoses,
AUTO
option is not available.
2.
Select the limit according to the hose with the ComWheel.
To produce a single measurement:
•
Press the
Start Cancel
side panel or module key, or
press the
NIBP
key and select
Start Manual
.
To measure automatically after set intervals:
•
Press the
Auto On/Off
side panel or module key, or
press the
NIBP
key and select
Start Cycling
.
To measure continuously for five minutes:
•
Press the
NIBP
key and select
Start STAT
.
During measurement
•
Observe the cuffed limb frequently. Measurement may impair
blood circulation. Intervals below 10 minutes and STAT
measurements are not recommended for extended periods of
time.
•
Make sure that tubes are not bent, pressed or stretched.
Measurement may be impaired.
•
Blood pressure values may be affected by a change in the
patient's position.
NOTE: When 30 minutes has passed from the latest NIBP
measurement, the numeric value turns gray.
NOTE: The presence of some arrhythmias during NIBP measurement
may increase the time required for the measurement. For details
about the test results of the functioning of the NIBP measurement in
the presence of arrhythmias, see the "User’s Reference Manual:
NIBP."
Stopping
To release the cuff pressure before the measurement is finished:
•
Press the
Start Cancel
side panel or module key, or
press
the
NIBP
key and select
Stop XX
.
Setting cycling intervals
1.
Press the
NIBP
key.
2. Select
Cycle Time
.
3. Select the interval time of the list with the ComWheel.
Using NIBP cuff for venous stasis
1.
Press the
NIBP
key.
2. Select
Start Ven.Stasis
.
Inflation limit
Venous stasis
pressure
Venous stasis
time
Infant
145 ±5 mmHg
30 ±5 mmHg
1 min
Child
200 ±10 mmHg
40 ±5 mmHg
2 min
Adult
280 ±10 mmHg
40 ±5 mmHg
2 min
WARNING: The monitor sets the inflation pressure
automatically according to the previous measurement.
Discharge the patient from the monitor to reset the
inflation limit before measuring a new patient.