MESSAGE
CAUSE
WHAT TO DO
NIBP manual
NIBP: Air leak or loose cuff have interrupted the
autocycling mode
Check setup and restart autocycling.
No battery backup
There are no batteries or they have both failed.
Replace the batteries; see section “Replacing the
Noise
ECG: Unreliable HR calculation or distorted waveform,
may occur during defibrillation or because of motion
artifacts.
Check the patient status. Check the electrodes.
No SpO2 probe
SpO
2
:
The probe is not properly connected to the
module.
Check connection between sensor and module.
No SpO2 pulse
SpO
2
:
Pulse signal is poor.
Try other measuring sites.
No P1 transducer
InvBP: Invasive blood pressure transducer or channel x
cable disconnected.
Connect the transducer or the cable.
Poor signal
SpO
2
: Pulse signal is poor possibly because of low
perfusion.
Change measuring site.
'PSM' measurements removed
Measurement module has been removed.
Reconnect the module if you want to restart the
measurement.
RAM error
Memory checking failed.
Contact authorized service personnel.
Recorder module removed
The N-FREC or N-FCREC module has been removed.
Reconnect the module if you need a recorder.
Replace Battery A
Replace Battery B
There is hardly any charge left in one of the batteries.
Also check the symbol on screen.
Replace the battery as soon as possible; see
section “Replacing the batteries.”
Replace D-Fend
Gases: The water trap is partially blocked.
Replace the water trap.
Sample line blocked
Gases: The gas sampling line is blocked or the water
trap is occluded.
Change sampling line and water trap.
Select inflation limits
NIBP: You are using a hose without an automatic
identification. When you try to start the measurement,
the monitor goes automatically to the selection
NIBP
Setup
-
Inflation Limits
.
Select appropriate inflation limits. NOTE:
AUTO
option is not available for these hoses.
Sensor INOP
Gases: The gas measuring sensor is inoperative or the
temperature in the module has increased.
Contact authorized service personnel.