MESSAGE
CAUSE
WHAT TO DO
Check NIBP
NIBP measurement affected by low blood pressure and
pulsation, or a change in patient’s condition.
Check patient status, measurement setup and
cuff.
Check probe
SpO
2
:
There is no detectable SpO
2
signal, the sensor is
faulty or it is detached from the patient.
Check the sensor and connections.
Check sample gas out
Gases: Sample gas outlet is blocked.
Remove blockage.
Condition Battery A
Condition Battery B
Battery no longer holds its charge.
Condition the battery according to instructions of
the external charger.
Cuff loose
NIBP cuff is not attached to patient, it is too loose or the
hose is not connected to module.
Check cuff and hose.
Cuff occlusion
NIBP tubes or hose kinked.
Check tubes and hose.
Cuff overpressure
NIBP cuff is squeezed during measurement.
Check NIBP cuff, hose and tubes. Restart.
EEPROM error or EPROM error
Memory checking failed.
Contact authorized service personnel.
Gas measurements removed
Gases: The N-FREC or N-FCREC module has been
removed.
Reconnect the module if you want to restart the
CO
2
measurement.
Identical modules
Another module measuring the same parameter is
already installed in the system.
Remove the other module. Note that you can only
use one E-PSM(P) and/or one N-Fx module at a
time.
Infl. limits! Check setup
NIBP: Adult or child cuff is used but infant inflation limit
is selected.
Check cuff and inflation limits.
InvBP's not zeroed
InvBP: One or both channels have not been zeroed.
Zero the channel indicated or zero both channels.
Leads off
ECG trunk cable, all leadwires or neutral electrode
(RL/N) are disconnected. Offset voltage between two
electrodes is too high.
Reconnect the disconnected electrode or leadwire.
Low perfusion
SpO
2
: Signal is weak possibly because of low perfusion. Change measuring site.
MVexp << MVinsp
Exhaled volume is markedly smaller than inhaled.
Check the whole system for leakages.
Network down: xxx
(XXX = network name)
Network cable is not connected.
If the monitor is used with WLAN option, it is in shadow
region and not connected to the network.
S/5 iCentral is shut down.
Check the cable.
Check signal strength. If it is ok, check the
iCentral.
Check the iCentral.