Connecting external devices
1
2
3
4
5
(1) Label specifying the external device
(2) LED
indicators
(3) Black bus cable to/from another
interfacing module if needed
(4) Gray device specific cable: connect
to the communication port of the
external device
(5)
Black bus cable to the monitor’s
connector for DIS (or to another
interfacing module)
WARNING: Make sure that
you are connecting the
interfacing module to the
device specified in the label.
•
Turn off the external device you want to interface.
•
Connect the gray device specific cable to the external device and
the black bus cable to the monitor’s connector for the S/5 Device
Interfacing Solution, see page 9, or to another interfacing module.
•
Turn on the external device.
NOTE: The status message ‘Connected’ appears on the
Status Page
of the
Interfacing
menu after you have connected the external device
to the interfacing module and turned it on. Note also that the monitor
and the interfacing module must be operational.
CAUTION: Make sure that the interfacing module is always
used in vertical position to prevent water from entering the
module.
Checking the functioning of the S/5 DIS
You can check the functioning of the S/5 DIS in two ways:
1. Press the
Monitor Setup
key and select
Interfacing
-
Status
Page
. The
Status Page
shows you the current communication
status of the interfacing modules connected to the bus (1 - 4 pcs).
2. Check the LED indicators on the interfacing module:
GREEN YELLOW INDICATION
lit
dark
Physical connections between the monitor, interfacing
module and external device are in order. The device has
been selected in the menu.
dark
lit
There is something wrong with the physical connections
between the monitor, interfacing module and external
device. The external device has not been selected in the
menu, see the “User’s Reference Manual.”
lit
lit
Physical connections between the monitor, interfacing
module and external device are in order but the external
device has not been selected in the menu; see the “User’s
Reference Manual.”
dark
dark
The interfacing module is not connected to the monitor
Selecting the external device
•
Press the
Monitor Setup
key and select
Interfacing.
•
Select the desired measurement parameter group (like SvO
2
).
•
Select the connected external device by its name.
NOTE: The device appears on the list only if it is correctly connected.
Summary of Contents for S/5 FM
Page 1: ...S 5 FM User s Guide guide_FM_uusi pmd 03 04 2004 8 31 1 ...
Page 22: ...20 20 For your notes ...
Page 32: ...30 30 For your notes ...
Page 60: ...58 58 For your notes ...