![Datex-Ohmeda S/5 FM User Manual Download Page 44](http://html.mh-extra.com/html/datex-ohmeda/s-5-fm/s-5-fm_user-manual_513840044.webp)
42
42
Recording on alarms
1. Press the
Print/Record
key
.
2. Select
Record Waveforms
.
3. Select
Start on Alarms - YES
.
Recording is activated when the following alarms reach the red level:
Asystole, Tachy/Brady, Art High/Art Low, V Fib, V Tachy; V Run >3 with
Extended arrhythmia analysis.
Art and ECG1 waveforms are recorded.
Selections are preconfigured.
Recording calculations
1. Press the
Wedge C.O.
key.
2. Select
Calculations
.
3. Select
Hemodynamic Calcs
or
Oxygenation Calcs.
4. Select
Record Calcs
.
Recording waveforms
You can record three waveforms to a local recorder, and two to four
waveforms to a network recorder:
1. Press the
Print/Record
monitor key.
2. Select
Record Waveforms - Record Wave
.
3. If the monitor is connected to the network, you can also use the
network recorder by selecting
Record to Net
. The network
recorder uses the settings of the S/5 iCentral.
4. Stop recording by selecting
Stop Wave
.
Changing the paper speed
To see the waveforms more clearly or more generally, change the
paper speed:
1. Press the
Print/Record
key.
2. Select
Record Waveforms - Paper Speed
.
To select other waveforms for recording, press the
Print/Record
key
and select
Record Waveforms - Waveform 1, 2,
or
3
.