4a Tests and Calibration 4.X Software
4a-6
05/04 1006-0453-000
4a.3.1 Alarm Log
The Alarm Log displays up to 10 of the most current alarm messages that have
been logged. Each log entry shows:
•
Bootup Count
- number (the bootup count is incremented each time the
machine is turned on).
•
Time (ms)
- the time in milliseconds since bootup when the latest alarm
condition occurred.
•
# Times
- the number of times that the specific alarm condition has occurred
during the noted bootup count.
•
Alarm
- message associated with the particular alarm condition.
Figure 4-3 • Alarm Log menu
The bottom left corner of the screen displays additional information:
• Bootup Count Last Cleared
• Current Bootup Count
Clear Alarm Log
Select to clear the alarm log. The system will ask you to confirm that you want
to clear the log.
Remarks
After the Alarm Log is cleared:
• the Bootup Count Last Cleared number will be reset to the Current Bootup
Count number.
• the menu will show the message “No entries in log!”.
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......