5 Troubleshooting
5-14
05/04 1006-0453-000
High O2
medium
In-range O
2
data
available and alarm
limit not = off
O
2
> high limit
Check O
2
limit. Is actual O
2
higher than limit
setting?
Verify O
2
calibration has been done correctly.
Does calibration pass?
If calibration does not pass, replace O
2
cell.
If there is still a calibration problem, replace
the SIB board.
Alarm condition becomes false or enabling criteria
becomes false.
High Paw
(2)
high
(5)
In-range P
aw
data
available
P
aw
> P
limit
Check Pressure limit setting is not set too low
for ventilator settings and breathing circuit.
Check the breathing circuit for occlusions.
Perform flow sensor calibration.
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Alarm condition becomes false for 15 seconds or
until the end of the next full inspiratory cycle.
High V
E
medium
In-range flow data
available and alarm
limit not = off
V
E
> high limit
Check ventilator settings and volume output.
Check the breathing circuit and flow sensor
connections.
Perform flow sensor calibration.
Replace the flow sensors.
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Alarm condition becomes false or volume alarms
state transition to standby (or off).
High V
TE
medium
In-range flow data
available and alarm
limit not = off
V
TE
> high limit
Check ventilator settings and volume output.
Check the breathing circuit and flow sensor
connections.
Perform flow sensor calibration.
Replace the flow sensors.
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Alarm condition becomes false or volume alarms
state transition to standby (or off).
Insp Flow
Sensor Fail
User Alarm
The system cannot read
the calibration data
stored in the sensor.
Operation continues
with default values.
Replace the flow sensor.
Insp Reverse
Flow
(5)
medium
(1)
In-range flow data
available in mechanical
ventilation
Negative flow on
inspiratory sensor
during expiration in
circle system
Check flow sensor connections for “No Flow
Sensor” alarm.
Check the breathing circuit
Perform flow sensor calibration.
Check insp/exp check valves.
Replace the flow sensors.
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Two consecutive breaths measured with “correct”
flow seen by flow sensors
Message
Alarm type/
Priority
Enabling Criteria
Alarm Condition
Service Repair
Removal Criteria
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......