![Datex-Ohmeda Aestiva 7900 SmartVent Technical Reference Manual Download Page 121](http://html.mh-extra.com/html/datex-ohmeda/aestiva-7900-smartvent/aestiva-7900-smartvent_technical-reference-manual_513817121.webp)
5 Troubleshooting
1006-0453-000 05/04
5-5
10049
Minimum system shutdown:
display voltage test failed
Reboot system. If problem continues,
For machines with original CPU:
1. Measure voltage at TP 22 of the PSB (Section 6.6.1).
Verify VH_EL = +14.5V (10-15V).
If in spec, replace PSB to CPU cable; otherwise continue.
2. Disconnect the Display cable. Measure again.
If in spec, replace EL Display; otherwise continue.
3. Disconnect CPU board. Measure again.
If in spec, replace CPU board; otherwise replace PSB.
For machines with Integrated CPU:
1. Measure voltage at TP 213 of the CPU (Section 6.6.3).
Verify VH_EL = +14.5V (10-15V).
If in spec, replace CPU; otherwise continue.
2. Disconnect the Display cable. Measure again.
If in spec, replace EL Display; otherwise replace CPU.
1004A
Minimum system shutdown:
A/D converter failure
Reboot system. If problem continues, replace CPU board.
1004B
Minimum system shutdown:
CPU test failure
Reboot system. If problem continues, replace CPU board.
1004C
Minimum system shutdown:
flash EPROM failure
Reboot system. If problem continues, replace CPU board.
1004D
Minimum system shutdown:
external RAM failure
Reboot system. If problem continues, replace CPU board.
1004E
Minimum system shutdown:
display RAM failure
Reboot system. If problem continues, replace CPU board.
1004F
Minimum system shutdown:
software watchdog failure
Reboot system. If problem continues, replace CPU board.
10050
Minimum system shutdown:
sequential watchdog failure
Reboot system. If problem continues, replace CPU board.
10051
Minimum system shutdown:
CPU clock too fast
Reboot system. If problem continues, replace CPU board.
10052
Minimum system shutdown:
CPU clock too slow
Reboot system. If problem continues, replace CPU board.
10053 through
10057
Minimum system shutdown:
CPU internal error
Reboot system. If problem continues, replace CPU board.
10058
Minimum system shutdown:
V_AUX test failed
Reboot system. If problem continues,
For machines with original CPU:
1. Measure voltage at TP 19 of the PSB (Section6.6.1).
Verify V_AUX = +4.988 to +6.09V.
If in spec, replace PSB to CPU cable; otherwise continue.
2. Disconnect CPU board. Measure again.
If in spec, replace CPU board; otherwise replace PSB.
For machines with Integrated CPU:
1. Replace CPU board
10059
Minimum system shutdown:
external reference voltage failed
Reboot system. If problem continues, replace CPU board.
1005A
Minimum system shutdown:
bootup memory failure
Reboot system. If problem continues, replace CPU board.
Error code (hex)
Description of error logged
Service action
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......