7 Repair Procedures
1006-0453-000 05/04
7-19
Figure 7-17 • Gas inlet valve
To service the gas inlet valve (GIV):
Note: You can service the gas inlet valve without removing it from the Vent
Engine.
1. Remove the tube guide plate from the top of the GIV.
2. Remove retaining ring and GIV cap.
3. Remove the shuttle and replace the lower O-ring and the lower U-cup seal.
Lubricate the seal lightly with KRYTOX.
Note
Use pneumatic pressure to remove the shuttle. Cover the shuttle with a soft
cloth and apply pressure (connect the drive gas hose or use pipeline pressure)
through the drive gas inlet.
4. Reinstall the shuttle and upper U-cup seal.
5. Reinstall and lubricate upper O-ring lightly with KRYTOX.
6. Reinstall inlet valve cap.
7. Install the retaining ring flat side out (away from the valve).
8. Reassemble in reverse order.
9. Perform the Checkout Procedure found in Section 3 of the Aestiva
Anesthesia Machine Service Manual.
4
5
6
1
2
3
4
7
8
AB.23.150
AB.23.149
AB.23.148
1. Retaining ring
2. GIV cap
3. O-ring, upper
4. Upper U-cup seal
5. Shuttle
6. Lower U-cup seal
7. O-ring, lower
8. Apply KRYTOX
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......