5 Troubleshooting
5-18
05/04 1006-0453-000
O
2
sensor
out of circ
User Alarm
No breathing circuit
module installed.
Install a breathing circuit module and an O
2
sensor.
On Battery -
Power OK?
User Alarm
The mains supply is not
connected or has failed
and the system is using
battery power.
Ventilate manually to
save power. At full
charge, the battery
permits approx. 30 min
of mechanical
ventilation.
Make sure power is connected and circuit
breaker is closed. Check ventilator fuse.
Patient
Circuit Leak?
User Alarm
Exhaled volume <50%
of inspired volume for at
least 30 seconds
(mechanical
ventilation).
Check breathing circuit and flow sensor
connections.
Paw < -10
cm H
2
O
high
(5)
In-range Paw data
available
Paw < -10 cm H
2
O
Check the breathing circuit.
Perform flow sensor calibration (10).
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Alarm condition becomes false.
PEEP Not
Achieved
low
In-range Paw data
available
Pmin does not reach
within 2
cm H
2
O
of
PEEP by end of
mechanical expiration
for 6 breaths in a row.
Check settings for not enough expiration time.
Check the breathing circuit.
Occlusions
Check scavenging
Check ventilator output with PEEP off
Perform flow sensor calibration.
Check that the patient airway pressure display
matches the airway pressure gage. If they
match, the problem may be pneumatic,
(supply pressure, regulator, inlet or flow valve,
etc.).
Check SIB to CPU cabling.
Replace the SIB board.
Replace the CPU board.
Alarm condition becomes false or any control
setting changed (except Plimit) or mechanical
ventilation transition on to off.
Message
Alarm type/
Priority
Enabling Criteria
Alarm Condition
Service Repair
Removal Criteria
Summary of Contents for Aestiva 7900 SmartVent
Page 1: ...Aestiva 5 7900 Anesthesia Ventilator Technical Reference Manual ...
Page 44: ...Notes 2 28 05 04 1006 0453 000 ...
Page 82: ...Notes 4a 36 05 04 1006 0453 000 ...
Page 116: ...Notes 4b 34 05 04 1006 0453 000 ...
Page 150: ...Notes 5 34 05 04 1006 0453 000 ...
Page 158: ...Notes 6 8 05 04 1006 0453 000 ...
Page 182: ...Notes 7 24 05 04 1006 0453 000 ...
Page 202: ...Notes 8 20 05 04 1006 0453 000 ...
Page 203: ......