DT1400 xx xx
20170920-V0
49
7.14. Clear minimum and maximum values
(15. menu item)
Function
To clear the measured and stored mimimum and maximum
values.
Sequence of operations
1. Enter the menu by typing the supervisor code.
The way the code should be type in is described in Chapter
7.1. Typing the code in.
You see this on the display:
.
2. Keep stepping by pressing the
▲
button as long as you
see this menu item:
(
R
eset
M
inimum-
M
aximum =
clear minimum-maximum value).
3. Enter the menu item by pressing the
ENTER
button.
4. The mnemonic
is shown on the display. The
mnemonic is blinking.
5. Now you still have the possibility to exit. If you want to
exit, as you do not want to clear the minimum and maximum
values after all, press the
MENU
button.
6. If you want to clear the minimum and maximum values,
press the
▲
button to change the „NO” to „YES”.
•
= I do not want to clear
•
= I want to clear
7. Press
ENTER
button you will see on the display:
(
Ready
) the clear has done.
Exit from the menu item
1. After you have performed the setting, press the
MENU
button. This takes you out from the menu item, and you see
this:
.
(2. If you do not want to exit from the menu because you
want to perform other settings too, you may press the
▲
button to select the desired menu item.)
Exit from the menu
Press the
MENU
button. First the mnemonic
(
Save
),
then the mnemonic
(
Exit
) are shown on the display.
With this the storing of the settings is completed. The
instrument has exited the menu and goes on with the
measurement.
Summary of Contents for DT1400 RL2 IA
Page 1: ...DT1400 xx xx Limit switch Operating Instructions ...
Page 68: ...DT1400 xx xx 68 20170920 V0 10 2 Application example Analog output option ...
Page 69: ...DT1400 xx xx 20170920 V0 69 RL4 option ...
Page 70: ...DT1400 xx xx 70 20170920 V0 TS option ...
Page 71: ...DT1400 xx xx 20170920 V0 71 ...
Page 72: ......