Scope marker view
Items
Description
1
m1/m2 points
Shows where the measurement markers intersect the selected waveform. Data values at the bottom of the dis-
play are referenced to the intersect locations of the markers. Calculations are based on the data points in
between the intersect locations.
2
Delta
Indicates the delta or absolute difference between the markers in units (volts, amps, or watts) and in time
(seconds). The value in parenthesis is the frequency, which is the reciprocal of the time (1/time).
3
m2
Indicates the m2 marker value in volts, amps, or watts at the intersection point. Also indicates the distance in
time that the m2 marker is in relation to the present trigger position. If a marker is out of view, an arrow indic-
ates the direction of the marker
.
4
m1
Indicates the m1 marker value in volts, amps, or watts at the intersection point. Also indicates the distance in
time of the m1 marker in relation to the present trigger position. If a marker is out of view, an arrow indicates
the direction of the marker
.
5
Min
Indicates the minimum data value (in volts, amps, or watts) between the marker locations of the selected wave-
form. Also indicates the distance in time of the minimum value in relation to the present trigger position.
6
Max
Indicates the maximum data value (in volts, amps, or watts) between the marker locations of the selected
waveform. Also indicates the distance in time of the maximum value in relation to the present trigger position.
7
Avg. (if selected)
Calculates the average data value (in volts, amps, or watts) between the marker locations of the selected wave-
form. Time indicates the time between markers over which the average value is calculated.
8
RMS (if selected)
Calculates the rms value between the marker locations.
9
Vp-p (if selected)
Calculates the difference between the maximum and minimum values. Time information is not valid for
calculated p-p values.
Keysight E36150 Series User's Guide
135