Set the trigger source to Key/Immediate:
TRIG:SOUR IMM
To enable the output:
OUTP ON
To initiate the transient trigger system:
INIT:TRAN
Ramp Arb (Option E36150ADVU)
Step 1– Select Arb waveform
Press Arb to access the Arb selection menu. Then select Ramp.
Press Back to return to Arb main menu.
Step 2 – Configure the Ramp Arbs
Configure the Ramp Arb accordingly. Refer to the below table for details.
Parameter
Available Key Settings
Description
Arb Type
Voltage or Current
Sets the output Arb type.
Default is Voltage.
Action required:
Press Arb Type to toggle between Volt and Curr.
Volt
Selects voltage as output Arb type.
Curr
Selects current as output Arb type.
Start Setting (I0 or
V0)
minimum to maximum
Sets the setting before the ramp.
End Setting (I1 or
V1)
minimum to maximum
Sets the setting after the ramp.
Delay (t0)
0 – 3600 s
Sets the delay time after the trigger is received but before the ramp starts.
Keysight E36150 Series User's Guide
87