To program step 0 as a voltage pulse:
ARB:SEQ:STEP:FUNC:SHAP PULS,0
ARB:SEQ:STEP:VOLT:PULS:STAR:TIM 0.25,0
ARB:SEQ:STEP:VOLT:PULS:TOP 10.0,0
ARB:SEQ:STEP:VOLT:PULS:TOP:TIM 0.5,0
ARB:SEQ:STEP:VOLT:PULS:END:TIM 0.25,0,(@1)
To program step 1 as a voltage ramp:
ARB:SEQ:STEP:FUNC:SHAP RAMP,1
ARB:SEQ:STEP:VOLT:RAMP:STAR:TIM 0.25,1
ARB:SEQ:STEP:VOLT:RAMP:END 10.0,1
ARB:SEQ:STEP:VOLT:RAMP:RTIM 0.5,1
ARB:SEQ:STEP:VOLT:RAMP:END:TIM 0.25,1
To program step 2 as a voltage sine wave:
ARB:SEQ:STEP:FUNC:SHAP SIN,2
ARB:SEQ:STEP:VOLT:SIN:FREQ 1,2
ARB:SEQ:STEP:VOLT:SIN:OFFS 10.0,2
ARB:SEQ:STEP:VOLT:SIN:AMPL 5.0,2
To repeat step 0 twice:
ARB:SEQ:STEP:COUN 2,0
To end the sequence at the last Arb value:
ARB:SEQ:TERM:LAST ON
To make the entire sequence repeat twice:
ARB:SEQ:COUN 2
To set the voltage mode to Arb:
VOLT:MODE ARB
To set the current mode to Arb:
CURR:MODE ARB
To set the trigger source to Key/Immediate:
TRIG:SOUR IMM
To enable the output:
OUTP ON
To initiate and run the list:
INIT:TRAN
106
Keysight E36150 Series User's Guide