![Data Translation DT9800 Series User Manual Download Page 94](http://html.mh-extra.com/html/data-translation/dt9800-series/dt9800-series_user-manual_511240094.webp)
Chapter 4
82
Continuous A/D Operations (cont.)
Set up buffering (see
).
Configure the subsystem using
olDaConfig
.
1
Deal with messages and buffers
(see
).
Stop the operation (see
Clean up the operation (see
Start the operation with
olDaStart
.
Continued from previous page.
Note:
To convert a voltage to temperature, linearize the voltage
for the specified thermocouple type, then subtract the CJC
temperature (10 mV per
°
C) from the linearized value. Refer to
the
Omega Complete Temperature Measurement Handbook and
Encyclopedia
for more information on linearizing values.
1
After configuration, if using an internal clock, you can use
olDaGetClockFrequency
to get the actual frequency that the internal pacer
clock could achieve; if using internal retrigger mode, you can use
olDaGetRetriggerFrequency
to get the actual frequency that the internal
retrigger clock could achieve.
Summary of Contents for DT9800 Series
Page 1: ...DT9800 Series UM 17473 L User s Manual ...
Page 4: ......
Page 13: ...1 1 Overview Features 2 Supported Software 7 Accessories 9 ...
Page 76: ...Chapter 2 64 ...
Page 77: ...65 3 Supported Device Driver Capabilities ...
Page 88: ...Chapter 3 76 ...
Page 112: ...Chapter 4 100 ...
Page 131: ...119 A Specifications ...
Page 143: ...131 B Connector Pin Assignments ...
Page 162: ...Index 150 ...