Chapter 5
110
8.
In the DAC1 Voltage box, click
0.000 V
.
9.
In the DAC1 Adjustment box, click the increment or decrement
arrows until your external meter display reads 0 V (within
0.005 V).
10.
In the DAC1 Voltage box, click
+9.375 V
and verify that your
external meter display reads +9.375 V (within ±12 mV.)
Note:
If you are not satisfied with the analog output calibration,
you can load the factory default settings stored in the EEPROM by
clicking
Restore
in the D/A Calibration Factory Settings box.
Once you have finished this procedure, the analog output circuitry is
calibrated.
To close the Calibration Utility, click
Done
.