background image

Micromon

Micromon

RS.8A.Y3.02 

          

©

Danfoss

01-2003

9

EKS

EKS111 probe (-35 to +85°C)

Digital

NO

Normally open

NC

Normally closed

To set the units of measurement, repeatedly press ‘

↵↵

↵↵

’ to scroll through the fields

until the Units field is highlighted.

Enter the units required using the ‘

’ or ‘

’ keys to scroll through the character

list and the’

' or ‘

’ key to move the cursor left or right respectively.

Note: This is a text field only and does not effect the operation of the system.

You can use the ‘

‘ and ‘

’ keys together to delete the character before the

cursor and ‘

‘ to change case.  When the units have been set, press ‘

↵↵

↵↵

’.  The

edit box will move onto the next field.

Select the point as previously described. Repeatedly press ‘

↵↵

↵↵

’ to scroll through

the fields until the High field is highlighted.  Set this field to set the value at or
above which an alarm will occur by using the ‘

’ or ‘

’ keys to scroll through the

character list and the ‘

’ or ‘

’ key to move the cursor left or right respectively.

When the high temperature alarm point has been set, press ‘

↵↵

↵↵

’.  The edit box will

move onto the next field.
Repeat the operation for the Low field, to set the value at or below which an alarm
will occur.  Press ‘

↵↵

↵↵

’ to move the edit box onto the next field.

Repeat the operation for the Delay field.  Set the length of time (in minutes) that the
fault must be present before an alarm is raised.

Log printouts are available for selected points on a timed or demand basis via the
print menu. 'Timed' means the Micromon is programmed to automatically print a

log at preselected intervals. A log printout can also be initiated via the ‘

‘ key on

the keypad (manual). When a log is printed only points with the log action field
appropriately set will be output. The available settings are:

Off

Set this point to OFF if you do not want this point to be
shown on any log printouts.

Timed

The point will be included only in a 'Timed' log printout.

Manual

The point will be included only in a 'Manual' log printout.

On

Set this option to ON if you require this point to be included in all

log printouts.

Repeatedly press ‘

↵↵

↵↵

’ to scroll through the fields until the Log field is highlighted.

Use the '

' or '

' keys to set the required log action and press ‘

↵↵

↵↵

’ to move the

cursor onto the next field.

Graph printouts are available for selected points on a timed or demand basis via
the print menu. 'Timed' means the Micromon is programmed to automatically print

a graph at preselected intervals. A graph printout can also be initiated via the ‘

key on the keypad (manual).When a graph is printed only points with the graph
action field appropriately set will be output. The available settings are:

Off

Set this option to OFF if you do not want this point

graph to be shown on any graph printouts.

Timed

The graph for this point will be included only in a 'Timed' graph

printout.

Manual

The graph for the point will be included only in a 'Manual' graph

printout.

Alarm Settings

Log Action

Graph Action

Printout Options

Units of Measurement

Summary of Contents for Micromon

Page 1: ...Monitoring and Alarm System Micromon Manual Refrigeration and Air Conditioning Controls R E F R I G E R A T I O N A N D A I R C O N D I T I O N I N G...

Page 2: ...PointType 10 Passcode Setup 10 Log Graph Setup 10 Modem Dialout Setup 11 Dialout 11 CallType 11 Number 12 Pager Info 12 Send 12 Modem Configuration 12 Baud Rate 12 Point Detail 13 Point Overview 13 O...

Page 3: ...inted on a regular basis The data is accessible both on and off site via the external modem connection and alarm calls can also be gener ated if required The Micromon has been designed in such a way t...

Page 4: ...nsors The Micromon provides audio and visual alarms whenever an input exceeds its pre set alarm levels The system can be pre set to dial out using a digital dialler or via a modem whenever an alarm oc...

Page 5: ...ey will force the highlight to the bottom of the menu In addition when a character entry field has been selected this key is used to increment the displayed value in the entry field SecondaryFunctions...

Page 6: ...ursor to the Graph screen and selects points on the Overview screen Backspace in edit fields Adjusts Screen Contrast Accented special characters editing mode or Sort and display alarms by point when i...

Page 7: ...nt up or down to the required number This screen is used select the application of those points that have just been selected in the previous screen Use the and keys to move the cursor to the required...

Page 8: ...y to move the active field box around the temperature units field then use the or key to toggle betweenFahrenheitandCentigrade The date format of the Micromon unit can be set to either DD MM or MM DD...

Page 9: ...selected points on a timed or demand basis via the print menu Timed means the Micromon is programmed to automatically print a log at preselected intervals A log printout can also be initiated via the...

Page 10: ...he Passcode Setup option Use the to highlight the passcode to be modified Type in the passcode using the or keys to scroll through the character list and the or key to move the cursor left or right re...

Page 11: ...this field to specify the type of device which is connected to the specified telephone number The options are as follows NO DIAL Indicates the Micromon will not dialout GSM SMS Indicates that a RS232...

Page 12: ...s S Sends the Store Name A Sends the Last Alarm Message C Sends a count of alarms and muted alarms in the format2 4 R Sends the Carriage Return CR character L Sends the Line Feed LF character Sends th...

Page 13: ...the screen To display more details of a particular point position the cursor box around the point required and press The Point Detail screen will be displayed Note If a point has a local offset appli...

Page 14: ...sses etc These offsets are point specific and may only be defined from the Micromon keypad although their value is readable when accessing the unit remotely Under normal circumstances these offsets if...

Page 15: ...te Options which have a symbol alongside them are password protected Use this option to display the default Point Overview screen as previously described To display the alarm history select the Alarm...

Page 16: ...ired value and the key to move through each field The Printing In Operation screen is then displayed and the data is sent to the printer Each alarm is given a number from 1 to 999 restarting at 1 afte...

Page 17: ...e n h c a e n o e v i t c A d e s s e r p s i y e k E T U M m r a l A p e R 2 Y A L E R a r o f n o i t c e n n o c d e y a l e R t i n u m r a l A t a e p e R l l i t s e r a s m r a l a e l i h w e...

Page 18: ...viced only by qualified service personnel and should ensure that replacement parts match the original safety characteristics CAUTION This unit is fitted with rechargeable batteries use only manufactur...

Page 19: ...M can be used to control the modem power The Relay is rated at 24V 1A max so the low voltage power after the mains transformer must be switched Mains voltage MUST NOT be connected to this relay Modem...

Page 20: ...that such alternations can be made without subsequential changes being necessary in specifications already agreed All trademarks in this material are property of the respecitve companies Danfoss and...

Reviews: