SALC-212-174
42
7.3 Test
Test mode is selected by pressing SET and MODE switches
continuously for longer than 3 seconds.
Pushing MODE switch performs movement of setting item.
The present mode can be returned to the display mode by
pressing DISPLAY switch.
(1) Alarm output check 【 With an option 】
Even if this product does not have input, it can perform ON/OFF test of an alarm output (relay-contact output).
Whenever it pushes SET , ON and OFF change.
Default:OFF
Output OFF Output ON
Alarm test
(2) Analog output check 【With an option】
Even if this product does not have input, it can test analog output (three circuits).
Ao1 check
RESET/
SHIFT
MAX/MIN
RESET/
SHIFT
MAX/MIN
Ao2 check
Ao3 check
RESET/
SHIFT
MAX/MIN
◆ AO (analog output) 1 to 3 check.
Sets the output factor about each analog output.
It selects 0%(4mA), 50%(12mA), and 100%(20mA)
with + or - switch. If SET is pushed,
analog output will output.
Default:0% (4mA)
Display
mode
SET and MODE
3 seconds
DISPLAY
Alarm output check
MODE
MODE
DISPLAY
Analog output check
MODE
DISPLAY
LCD display check
SET
The status of lights
at the output time.
The status of
lights at the
output time.
Analog output
factor
Output value