background image

SALC-212-174 

 

38

 
 

6.2  Performance. 

Item 

Specification 

Intrinsic error 

Reference to measure specification and intrinsic error 

Accuracy of bar graph  ±10% (% for span) 
Influence by 
temperature 

Within accuracy by 23±10℃. 

Compliance standard 

JIS C 1102-1:2007 ,JIS C 1102-2, -7, -9:1997 , JIS C 1111:2006  

Safety 

JIS C 1010-1:2005 
CATⅢ (The category to the measurement performed with fabric equipment)   
 Maximum circuit voltage 300V 
Pollution degree 2 (Usually, environment which only contamination of non-conductivity 

generates.  However, temporary conductivity which originates in  
dew condensation depending on the case occurs.) 

Display updating time  About 1 second (Bar graph:0.25 seconds)  

Display device 
Display composition 

LCD 
 Number,  Character, 
 Segment color:Black   

Main monitor 

4 digit, character height 11mm 

Sub monitor (Left) 

4 digit, character height  6mm 

Sub monitor (Center),(Right)  4 digit, character height  6mm 
Bar graph 

20 dots  

LCD view angle  

Standard 

For upper installation  
(For lower view) 

Upper view angle 10°,Lower view angle 60°, 
Right and left view angle 60° 

Special 

For lower installation  
(For upper view) 

Upper view angle 60°,Lower view angle 10°, 
Right and left view angle 60° 

Backlight  

LED backlight:Green or White  
Always-on, Auto off (after 5 minutes without operating), Always-off.  Setting is possible. 
White backlight can select brightness from five steps of 1 to 5.  (

19

Auxiliary supply 

AC85~264V  50/60Hz  10VA (Rated voltage, AC100/110V,200/220V) 
DC80~143V            6W  (Rated voltage, DC100/110V) for both AC and DC uses 

Rush current  
(Time constant) 

Rated voltage AC110V   2.2A or less (About 2.5ms) 
Rated voltage AC220V   4.4A or less (About 2.5ms) 
Rated voltage DC110V   1.6A or less (About 2.5ms) 

Input consumption VA 

Current circuit 

0.1VA or less (5A) 

Overload capacity  

Current circuit 

40 times 1 second, 20 times 4 seconds, 10 times 16 seconds,  
1.2 times continuation of rated current. 

Auxiliary supply 

1.5 times 10 seconds, 1.2 times continuation of rated voltage. 
In case of DC110V, 1.5 times 10 seconds, 1.3 times continuation of  
rated voltage. 

Insulation resistance 

Between electric circuits and case (Earth).  

Above 50MΩ at DC500V megger 

Between input and output and auxiliary supply. 
Between analog output and alarm output. 
Between analog outputs are not insulation of minus common. 

Voltage test  
(Commercial frequency 
withstand voltage) 
JIS C 1102-1 
JIS C 1111 

Between electric circuits and case (Earth).  

AC2210V (50/60Hz) 5 seconds 

Between input and output and auxiliary supply. 

Between analog output and alarm output. 

Between analog outputs are not insulation of minus common. 

Impulse voltage test 
(Lightning impulse 
withstand voltage) 
JIS C 1111 

Between electric circuits and case (Earth). 
(An analog output is excluded) 

5kV 1.2/50μs  Positive and negative 
polarities, for each 3 time. 

Between input and auxiliary supply.  
(Grounds an output.) 

Note(

19

) About white backlight.  

The white backlight of this product is using white LED which combined the special phosphor and blue  
LED.  In the characteristics of this LED, color tone may be different for each product. 

 

Summary of Contents for SALC-110L

Page 1: ...SALC 212 174 INSTRUCTION MANUAL POWER LINE MULTI METER AMMETER SALC 110L 1φ2W 1φ3W 3φ3W ANALOG OUTPUT HARDWARE MODEL C Green backlight HARDWARE MODEL F White backlight ...

Page 2: ... We are also not responsible for damage caused by false operation in combination with connecting equipment which is beyond our control Do not disassemble remodel and repair this product Have a contact with us or sales agent when product failure happens to prevent fire electric shock injury Do not get this product wet to prevent heat generation ignition product failure When this product gets wet st...

Page 3: ... this product may become hard to see And deformation of a case may take place by the surface temperature rise The warranty period and warranty scope Warranty period The warranty period of this product is for one year after supplying the appointed place Warranty scope In the state of the normal use of product specification within the range according to this instruction manual the trouble within the...

Page 4: ...it at a dusty place Don t mount the unit at a place exposed to direct sunlight Avoid exposing the unit to direct sunlight even through a glass window If the meter is directly exposed to sunlight the surface temperature of the meter rises and the case may be deformed if the temperature exceeds 80 If the average temperature around the meter exceeds 40 the life of the unit may shorten 3 About dew con...

Page 5: ... current OR of each phase current Demand current OR of each phase current Reset method AUTO AUTO AUTO Contact delay time 0 second 0 second 0 second 3 Demand current detection Upper limit 80 0A 400 0A 40 00A Time interval 0 second 0 second 0 second 4 Backlight Action AUTO AUTO AUTO Brightness 4 3 Middle 3 Middle 3 Middle 5 Measurement range Current range 100 0A 500 0A 50 00A Current display intrins...

Page 6: ...n of our company otherwise the warranty does not apply to the unit any more Also modifications may cause troubles a fire or other accidents For specifications change etc please contact us 7 Maintenance and check Wipe off the surface of the unit with a dry soft cloth The liquid crystal display LCD may light during cleaning on the LCD face However this phenomenon is caused by the static electricity ...

Page 7: ...Setting display 15 4 3 Operation 16 4 3 1 The main monitor display factor change 16 4 3 2 Phase display change 16 4 3 3 Setting value check 17 4 3 4 Setting mode 18 4 3 5 Reset 19 5 Setting 5 1 Function table 21 5 2 Setting table 23 5 3 Setting in detail explanation 26 5 3 1 Setting mode 1 26 5 3 2 Setting mode 2 32 5 3 3 Setting mode 3 36 6 Specification 6 1 Specification and intrinsic error 37 6...

Page 8: ... uses The mounting method of this unit is compatible with the mounting method of conventional 110 square mechanical meter This unit is mounted at 2 diagonal points A backlight function is equipped Selection of backlight on backlight off and auto backlight off and setting of brightness only white backlight are possible Automatic turning off the lights at the time of non operation can be established...

Page 9: ...ue Upper limit setting index An upper limit set point is displayed Unit display It sets automatically by measurement range setting The switch to which measurement displays element of main monitor is changed If it is not operated for 10 minutes after a display change it returns to the original measurement display factor In set mode it is used as a switch that changes a setting value It is the switc...

Page 10: ...iagram Caution on handling Mount the LCD to obtain an optimum angle since the contrast changes according to the monitoring angle 1 For upper case installation 2 For lower case installation Installation 1 A product is put in a cut hole of a panel from a front A body is inserted until it exceeds retaining stopper of the lower base 2 Please fix a product certainly with attached M5 flange nut for inst...

Page 11: ...l for enhancing the shield effect Keep the grounding resistance between the grounding terminal and the ground to be lower than 100Ω 4 Keep a distance of more than 30cm between this unit and the circuit breaker as well as between this unit and the relay contact signal line 5 No protection is necessary for this unit even if the transmission line may be affected by an induced lightning surge or the l...

Page 12: ... terminates setting mode And it is used in case it returns the display combination of a measurement factor Convenient functions 1 In case a measurement change or a phase change is performed and the original screen composition is not clear anymore DISPLAY is pushed for more than 3 seconds or it s no operation for 10 minutes and returns to original screen structure 2 If a measurement change is perfo...

Page 13: ...measurement It displays the factor of the maximum and the minimum measurement on display on the main monitor MAX MIN Push 1 seconds RESET SHIFT Maximum Minimum reset All factor Push 1 seconds RESET SHIFT Phase display change Main monitor display element change Setting point check Or No operation 10 minute SET Setting mode 1 Push 3 seconds Or No operation 10 minute Setting mode 2 Push 3 seconds Or ...

Page 14: ... about the measurement factor which performs holding of the maximum value and the minimum value by switch operation These maximum values and the minimum value are reset by switch operation it updates to the instantaneous value at the time In addition as for the maximum value and the minimum value power supply reset is not cleared either And this display is held by even after 10 minutes of switch n...

Page 15: ...ethod is manual reset setting even if it returns from upper limit detection an alarm detection display is held The return in this case needs alarm reset operation Please refer to 4 3 5 Reset for alarm reset Alarm factor Example of a display Demand current Upper limit Detection display Alarm setting value At alarm factor setting Upper limit detection Demand current A V Maximum demand current A V A ...

Page 16: ...m output can be tested in this setting mode Setting mode 1 No 111 Display pattern Setting mode 2 Setting of measurement range analog output external operation input and measurement display ON OFF is performed And it can initialize settings in this setting mode Setting mode 2 No 211 Voltage range Setting mode 3 Setting of an input circuit is performed And analog output can be adjusted in this setti...

Page 17: ...Main monitor A U Main monitor A V Main monitor A W Default display 4 3 2 Phase display change 3 phase 3 wire Single phase 3 wire A phase display of current is changed Everything which is being indicated A change is performed by DISPLAY A measurement display and maximum display minimum display can also perform this operation In addition after changing a phase display if a switch is not operated for...

Page 18: ... is not operated for 10 seconds after a set point check it will return to the original measurement display automatically Note 8 It is not displayed if there is no option DISPLAY or No operation 10 seconds RESET SHIFT A V display A U display A W display Primary current Factor display Alarm factor return method Contact delay time 8 Demand current time interval Demand current upper limit value Settin...

Page 19: ...setting mode please refer to 5 Setting Setting mode 1 Press SET for longer than 3 seconds Setting mode 2 Press SET and RESET SHIFT together for longer than 3 seconds Setting mode 3 Press SET and DISPLAY together for longer than 3 seconds Reference It can operate by measurement display and maximum minimum display Setting mode 2 Setting factor Setting No SET 3 seconds SET RESET SHIFT 3 seconds SET D...

Page 20: ... continues and it has caused And this operation is unnecessary if an alarm return method is set as AUTO automatic return An output is also OFF to an alarm return In addition setting can perform same operation in external operation input Please refer to 5 3 2 Setting mode 2 3 external operation input setting for the setting method Please refer to 6 3 Option for the external operation input It conti...

Page 21: ...lue It resets all the maximum values and minimum values In addition setting can perform same operation in external operation input Please refer to 5 3 2 Setting mode 2 3 external operation input setting for the setting method Please refer to 6 3 Option for the external operation input By measurement display and the maximum and minimum measurement display it continues pushing RESET SHIFT and 1 seco...

Page 22: ...W A V 27 28 1φ3W A U 1φ2W A 121AL 9 Alarm factor Sets the output factor of alarm 3φ3W DA OR detection of each phase current 29 1φ3W DA OR detection of each phase current 1φ2W DA 122AL 9 Alarm reset method Sets the output action at the case of an alarm reset AUTO Automatic reset 29 123AL 9 Alarm contact delay time Sets the contact delay time of alarm 0 second 29 131H Demand current upper limit valu...

Page 23: ...ute input adequate to 0 5 or less in analog output OFF No operation 34 231 11 External operation input function Sets the function of the external operation input Alarm reset 35 241 Current ON OFF Sets the ON OFF of current measurement display ON 35 242 Demand current ON OFF Sets the ON OFF of demand current measurement display ON 35 251 Set value initialization Initializes the settings in the sett...

Page 24: ...t factor is entered Press DISPLAY Returns to display mode 34 Sets the output factor of analog output 2 AO2 222A Press SET and RESET SHIFT together for longer than 3 seconds Press MODE 211 221A Press RESET SHIFT Select an output factor by and Press SET 222A Selected output factor is entered Press DISPLAY Returns to display mode 34 Sets the output factor of analog output 3 AO3 223A Press SET and RES...

Page 25: ...play factor is entered Press DISPLAY Returns to display mode 27 28 Sets the display factor of bar graph 116 Press SET for longer than 3 seconds Press RESET SHIFT Press RESET SHIFT 111 112 Press RESET SHIFT Press RESET SHIFT Press RESET SHIFT 113 114 115 116 Select an display factor by and If a sub monitor is selected an underbar will be displayed on the bottom of a digital display Press SET Select...

Page 26: ...tput Each parenthesized number shows a setting number and this number is displayed on the setting screen Items Setting and operation procedures Page Sets the reset method of alarm output 122AL Press SET for longer than 3 seconds Press MODE Press RESET SHIFT 111 121AL 122AL Select a reset method by and Press SET The selected reset method is entered Press DISPLAY Returns to display mode 29 Sets the ...

Page 27: ...3 seconds Press MODE 111 121AL Press MODE It is operation needlessness if there is no option Press MODE 131H 141 Press RESET SHIFT Select a brightness by and Press SET 142 Selected backlight brightness is entered Press DISPLAY Returns to display mode 31 5 3 Setting in detail explanation 5 3 1 Setting mode 1 Display mode SET 3 seconds 111 116 Display setting MODE 121AL 123AL Alarm output setting MO...

Page 28: ...U MDA U MDA DA U Displays set factor Single phase 3 wire Main monitor A U A W A N DA U DA W DA N MDA U MDA W MDA N Sub monitor Left A U A W A N Sub monitor Center A U A W A N DA U DA W DA N MDA U MDA W MDA N Sub monitor Right A U A W A N DA U DA W DA N MDA U MDA W MDA N Bar graph A U A W A N DA U DA W DA N MDA U MDA W MDA N Phase change Single phase 3 wire 14 Measurement factor change Measurement ...

Page 29: ... DA U 4 DA V 5 DA W 6 A V 2 A W 3 DA U 4 DA V 5 DA W 6 MDA U 7 MDA V 8 MDA W 9 Nothing 0 A U 1 MDA U 7 MDA V 8 MDA W 9 MDA U 7 MDA V 8 MDA W 9 111 Display pattern Select the factors to be measured and monitored by 4 digital displays out of combination patterns Set values are updated by SET 112 115 Main monitor Sub monitor left Sub monitor center Sub monitor right Sets these items for a display con...

Page 30: ...A W 3 phase 3 wire OFF Nothing 1 Demand current Single phase 122AL Alarm reset method setting Action at the case of a reset of alarm output can be selected from AUTO automatic reset and HOLD manual reset In AUTO automatic reset an alarm output is OFF according to a reset of an alarm In HOLD manual reset even after an alarm reset an output holds ON It performs a return output OFF in RESET SHIFT Sel...

Page 31: ...current DA Selection by and set value is updated by SET Default setting 0 second 0sec Nothing 5sec 10sec 20sec 30sec 40sec 50sec 1min 2min 3min 4min 5min 6min 7min 8min 9min 10min 15min 20min 25min 30min Demand time delay characteristic 0 20 40 60 80 100 0 10 20 30 Time min Demand value Time interval 10 minutes 95 Arithmetic method Demand current measurement Arithmetic method according with bimeta...

Page 32: ...automatically After that backlight will be turned on if either of switches is operated Selection by and set value is updated by SET Default setting AUTO Auto off 142 Backlight brightness white backlight products only It can select the brightness of backlight as five steps of 1 to 5 Backlight becomes the darkest if it is set as 1 Backlight becomes the brightest if it is set as 5 Selection by and se...

Page 33: ...AY DISPLAY DISPLAY DISPLAY Setting mode 2 is selected by pressing SET and RESET SHIFT switches continuously for longer than 3 seconds Pushing MODE switch performs movement of setting item The present mode can be returned to the display mode by pressing DISPLAY switch Caution If setting change should have been mistaken a display and output of measurement are not obtained correctly Therefore users m...

Page 34: ...of 100A is 120A A measurement range can be selected within the 40 to 120A Therefore can select a measurement range from the following 40 42 45 48 50 56 60 64 72 75 80 84 90 96 100 120A 1 1 0 1 2 1 4 1 5 1 6 1 8 2 2 0 2 4 2 5 2 8 3 3 0 3 2 3 6 4 4 0 4 2 4 5 4 8 5 5 0 5 6 6 6 0 6 4 7 7 2 7 5 8 8 0 8 4 9 9 0 9 6 213 Digit number of current range Sets the digit number of current range Selection by and...

Page 35: ...8 MDA V 7 MDA U 6 DA W 5 DA V At 3φ3W 224A Current output intrinsic sensitivity Output intrinsic sensitivity of rated input power value to an output upper limit value is set about each analog output of current The setting range can be selected from the following Setting range 40 0 120 0 0 1 step Selection by and set value is updated by SET Default setting 100 0 Current 225A Low input cut It is the...

Page 36: ... guidance and RESET are displayed by 7 segment display All reset ALL and RESET are displayed by turns by 7 segment display 4 241 242 Measurement ON OFF setting Measurement display ON OFF setting of each measurement factor is performed Selection by and set value is updated by SET Default setting ON All measurement factors 241 Current 242 Demand current RESET SHIFT MAX MIN RESET SHIFT MAX MIN 5 251 ...

Page 37: ... actual connection and phase wire are different 3φ3W 1φ3W U W N 1φ3W U V N 1φ3W V W N 1φ2W 2 321 326 Analog output adjustment With option The bias and span of each analog output are adjusted RESET SHIFT MAX MIN 321 Ao1 BIAS adjustment RESET SHIFT MAX MIN 322 Ao1 SPAN adjustment RESET SHIFT MAX MIN 323 Ao2 BIAS adjustment 326 Ao3 SPAN adjustment 325 Ao3 BIAS adjustment 324 Ao2 SPAN adjustment RESET...

Page 38: ...sible Main monitor Current Each phase Demand current Each phase Maximum demand current Each phase Sub monitor Left Current Each phase Sub monitor Center Current Each phase Demand current Each phase Maximum demand current Each phase Sub monitor Right Current Each phase Demand current Each phase Maximum demand current Each phase Bar graph Current Each phase Demand current Each phase Maximum demand c...

Page 39: ...tage AC100 110V 200 220V DC80 143V 6W Rated voltage DC100 110V for both AC and DC uses Rush current Time constant Rated voltage AC110V 2 2A or less About 2 5ms Rated voltage AC220V 4 4A or less About 2 5ms Rated voltage DC110V 1 6A or less About 2 5ms Input consumption VA Current circuit 0 1VA or less 5A Overload capacity Current circuit 40 times 1 second 20 times 4 seconds 10 times 16 seconds 1 2...

Page 40: ...eeds to be no malfunction 4 Electrostatic noise It is within 10 of errors in 8kV electrostatic noise at the case of an energization There needs to be no 10kV damage at the case of a non energization Condenser charge form Vibration JIS C 1102 1 Sweep vibration frequency range 10 55 10Hz Displacement amplitude 0 15mm Number of sweep 5 Sweep velocity 1 octave minute Shock JIS C 1102 1 Peak accelerati...

Page 41: ... to the maximum scale 1 step External operation input Function Three types of following functions can be operated by adding a voltage signal from the outside in addition to switch operation Alarm reset Alarm output is reset output OFF Please refer to 4 3 5 Reset about operation by the switch Maximum Minimum value reset The maximum minimum value is reset it updates to the instantaneous value at the...

Page 42: ... the wiring Outside the rated frequency 45 65Hz It cannot be used Cycle control SCR phase angle control PWM control or other inverter output is measured It cannot be used Analog output is not outputted Analog output is set to OFF or measurement factor is set to OFF Check the setting Alarm output does not return The return method is a manual reset Check the setting Settings changed It changed an in...

Page 43: ...tput ON Alarm test 2 Analog output check With an option Even if this product does not have input it can test analog output three circuits Ao1 check RESET SHIFT MAX MIN RESET SHIFT MAX MIN Ao2 check Ao3 check RESET SHIFT MAX MIN AO analog output 1 to 3 check Sets the output factor about each analog output It selects 0 4mA 50 12mA and 100 20mA with or switch If SET is pushed analog output will outpu...

Page 44: ...ay item may disappear Please wipe off with the dry cloth 7 5 Storage 1 Please avoid storage in the next space Low temperature high temperature high humidity and sunny place 2 The aluminum electrolytic capacitor is used for a product Please do the energization of the power supply within one year after shipment 7 6 Countermeasures against troubles As our principle we recall product in question and r...

Page 45: ...ce 11 13 Hitotsuya 1 chome Adachi ku Tokyo 121 8639 JAPAN TEL 81 3 3885 2411 Fax 81 3 3858 3966 Kyoto Office 1 19 Ichinobe Nishikawahara Jyoyou shi Kyoto 610 0114 JAPAN TEL 81 774 55 1391 Fax 81 774 54 1353 DATE November 22 2012 ...

Reviews: