standards. In case abnormal performance is observed, additional procedures, such as re-
orientation and replacement of the EAGLE Panoramic X-Ray Machine may be required.
b
It is recommended that field intensity should be lower than 3 V/m above the frequency range
of 150 kHz to 80 MHz.
Recommended separation distances between portable and mobile RF communication
equipment and the EAGLE Panoramic X-Ray Machine
The EAGLE Panoramic X-Ray Machine has been designed for use in electromagnetic
environments where RF radiated perturbations are controlled. The client or user of the EAGLE
Panoramic X-Ray Machine can help prevent electromagnetic interference by keeping a
minimum distance between portable and movable RF communication equipment (transmitters)
and the EAGLE S Panoramic X-Ray Machine, as recommended below, according to the
maximum output power of the communication equipment.
Maximum nominal
output power of the
transmitter (W)
Separation distance according to the transmitter’s frequency (m)
150 kHz up to 80 MHz
d = 1.2
√
P
80 MHz up to 800
MHz
d = 1.2
√
P
800 MHz up to 2.5 GHz
d = 2.3
√
P
0.01 0.12 0.12
0.23
0.1 0.38 0.38
0.73
1 1.2 1.2
2.3
10 3.8 3.8
7.3
100 12
12 23
For transmitters with maximum nominal output powers that are not listed above, the
recommended separation distance d in meters (m) can be determined by using the
equation applicable for the transmitter’s frequency, where P is the transmitter’s
maximum nominal output power in watts (W) according to the transmitter’s
manufacturer.
NOTE 1: In 80 MHz and 800MHz, the separation distance for the higher frequency range is
applied.
NOTA 2: These guidelines may not be applicable in all situations. Electromagnetic propagation
is affected by absorption and reflection by structures, objects and people.
WARNING
THE EQUIPMENT SHOULD NOT BE USED ADJACENT TO OR
STACKED ON OTHER EQUIPMENT, RECOMMENDATIONS
OF THIS MANUAL MUST BE FOLLOWED.
Summary of Contents for EAGLE
Page 1: ...USER S MANUAL...
Page 2: ...1 THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Page 7: ...6 2 SYMBOLS Use the icons below to identify the symbols on your equipment...
Page 65: ...64...
Page 66: ...65...
Page 67: ...66...
Page 74: ...73 16 12 CHARACTERISTIC CURVES OF THE X RAY TUBE Exposition Time...
Page 75: ...74 17 LABELS OF IDENTIFICATION 17 1 PACKAGE...
Page 76: ...75 17 2 PRODUCT...
Page 77: ...76 18 EQUIPMENT DIMENSIONS...