Medium patient selected
Large patient selected
kV
This function is used to fine
tune the kV after selecting the
patient age and size or to
directly select the kV.
The range of kV is from 60kV
to 85kV in increments of
2.5kV.
This value is selected using
PLUS/MINUS KEYS
If the kV is left unchanged the
equipment will be in
demonstration mode. In this
mode no x-ray is exposed.
No kV selected: Demonstration mode.
Example of kV selection: 75 kV.
mA
The anodic current is not user
adjustable. The value indicated
is optimum for image
generation in each profile.
Indication that current profile uses
8mA of anodic current.
TIME
In all profiles the value is not
user adjustable.
Value indicating that current profile
has 14 seconds of x-ray exposure.
IMAGE
LAYER
POSITION
This function allows the user to
adjust the image layer towards
the back or front of the dental
arch in panoramic profiles.
The adjustment is made using
the following keys:
-
PLUS
: moves layer
towards back of dental
arch.
-
MINUS:
moves layer
The image layer positioning change
commands are accepted when the
image layer icon is selected on the
screen.
Summary of Contents for EAGLE
Page 1: ...USER S MANUAL...
Page 2: ...1 THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Page 7: ...6 2 SYMBOLS Use the icons below to identify the symbols on your equipment...
Page 65: ...64...
Page 66: ...65...
Page 67: ...66...
Page 74: ...73 16 12 CHARACTERISTIC CURVES OF THE X RAY TUBE Exposition Time...
Page 75: ...74 17 LABELS OF IDENTIFICATION 17 1 PACKAGE...
Page 76: ...75 17 2 PRODUCT...
Page 77: ...76 18 EQUIPMENT DIMENSIONS...