8
3.4. OPERATIONAL ENVIRONMENTAL CONDITIONS
Operation ambient temperature range
+10 C to +35 C
(+50 F to 95 F)
Ambient temperature range recommended
+21 C a +26 C
(69.8 F to 78.8 F)
Operation relative humidity range
30% to 75%
(non condensing)
Atmospheric pressure range
700 hPa to 1060 hPa
(525 mmHg to 795 mmHg)
3.5. ADDITIONAL PROCEDURES PRIOR TO EQUIPMENT USE
Even prior to its first use, the equipment must be cleaned and disinfected; the same
additional procedures must be followed for reuse, as described in Chapter 11.
3.6. WARNINGS AND/OR CAUTION TO BE ADOPTED
3.6.1. WARNINGS AND/OR CAUTION DURING EQUIPMENT INSTALLATION
The equipment must be installed only by service technicians authorized by the
manufacturer.
Place the equipment on a site where it will not be in contact with moisture or
water.
Install the unit on a site where it will not be damaged by pressure, temperature,
moisture, direct sunlight, dust or salts.
The equipment must not be submitted to inclination, excessive vibration or shock
(including during transportation and handling).
This equipment has not been designed for use in facilities where vapors,
flammable anesthetic mixtures in contact with air, oxygen or nitrous oxide can be
detected.
Check the equipment’s voltage when performing electric installation.
Failure to
do so may damage the equipment.
The equipment must be properly grounded.
Failure to do so may result in
“Safety Hazard”.
Depending on local regulation the X-Ray emission control may require
installation outside the facility where the equipment is placed, and the operator
may need visual contact with the patient through a window with radiological glass
or similar, since the operator must not lose visual contact with the patient.
Mobile and portable RF communication equipment can affect the EAGLE
Panoramic X-Ray Machine.
This equipment must be solely used by health care professionals as it may cause
radio interference or interrupt the operation of nearby equipment. Mitigatory
measures, such as equipment re-orientation or replacement and the facility’s
screening, may be necessary.
Summary of Contents for EAGLE
Page 1: ...USER S MANUAL...
Page 2: ...1 THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Page 7: ...6 2 SYMBOLS Use the icons below to identify the symbols on your equipment...
Page 65: ...64...
Page 66: ...65...
Page 67: ...66...
Page 74: ...73 16 12 CHARACTERISTIC CURVES OF THE X RAY TUBE Exposition Time...
Page 75: ...74 17 LABELS OF IDENTIFICATION 17 1 PACKAGE...
Page 76: ...75 17 2 PRODUCT...
Page 77: ...76 18 EQUIPMENT DIMENSIONS...