62
Child – 10.5s – 8mA
Maxillary sinus
8s – 8mA
Open mouth + closed mouth TMJ
exposure time (TMJ 1 + TMJ 2 )
10s – 8mA
Bitewing
7.6s – 8mA
Mean magnification
1 : 1.22
Source to Image Distance SID
1.69 ft
Information Specific for Cephalometric Profiles
Analog cephalometric exposure
time/current
0,1 to 3s – 8mA
Complete digital cephalometric exposure
time/current
10 or 16,5 – 8mA
Low Dosage digital cephalometric
exposure time/current
6,6 or 11s – 8mA
Film dimension for analog cephalometric
exposure
8” x 10”
Mean magnification
1:1.1
Source to Image Distance SID
5.41 ft
For this equipment proper patient positioning is required to produce a good quality image.
The operator must stay away at least 9 feet from the equipment during exposure to
minimize the amount of ionizing radiation risk.
16.4. X-RAY GENERATOR
Generator type
High-frequency constant potential generator
Operating frequency
100 kHz
Maximum operation voltage
85 kVp
Heating and cooling curve
See Graphic on item 15.13 of this manual
Output power
680 W (85kV x 8mA)
Output power during 0,1s
680 W (85kV x 8mA)
Total filtration
0.01 ft Al eq. @ 85 kVp (This value takes all
mitigating circumstances that exist from the
emission source to the output of equipment)
Radiation escape
< 1.00mGy/h at 85kV / 8mA
Equipment
CLASS I - Type-B applied part
When submitted in charge the equipment
emits ionizing radiation.
Operation mode
Intermittent
The X-Ray Generator is mounted by the manufacturer.
X-Ray machine with radiologic protection according to NBR IEC 60601-1-3:2001.
X-Ray generator EAGLE NBR IEC 60601-2-7:1998
X-Radiation-emitting set EAGLE NBR IEC 60601-2-28:2001
Radiological equipment associated EAGLE NBRIEC60601-2-32:2001
16.5. TESTED EQUIPMENT LAW NORM
EN 60601-1 (1990);
Summary of Contents for EAGLE
Page 1: ...USER S MANUAL...
Page 2: ...1 THIS PAGE WAS INTENTIONALLY LEFT BLANK...
Page 7: ...6 2 SYMBOLS Use the icons below to identify the symbols on your equipment...
Page 65: ...64...
Page 66: ...65...
Page 67: ...66...
Page 74: ...73 16 12 CHARACTERISTIC CURVES OF THE X RAY TUBE Exposition Time...
Page 75: ...74 17 LABELS OF IDENTIFICATION 17 1 PACKAGE...
Page 76: ...75 17 2 PRODUCT...
Page 77: ...76 18 EQUIPMENT DIMENSIONS...