CU-ERT Operator’s Manual_ver 1.10
56
9. Maintenance
Although there are no user-serviceable parts inside the
Paramedic CU-ERT
, the user can perform
simple maintenance tasks to help prolong the life of the device and to help ensure that the device is in
mint condition. The following are the activities, together with frequency, that can be performed by the
user.
Used consumables should be disposed of in accordance with local regulations. Care should be taken in
disposing of spent external battery packs so as not to contribute to the pollution of the environment.
Maintenance Activities
Frequency
Activity
Actions to be Taken
Check supplies, accessories, and
spare parts for damage and
expiration.
If any supplies have expired, replace
them immediately.
Check the case of the
Paramedic CU-ERT
and the
accessories for any sign of
apparent damage.
If there is any apparent damage to the
case of the device, consult the
manufacturer.
Monthly and
after each use
Check for dirt Contamination.
If there is dirt contamination, clean the
case as suggested in Chapter 6 of this
Manual.
After each use.
Ensure that the connector of the
defibrillator pad assembly is
disconnected from the socket on
the
Paramedic CU-ERT
.
Disconnect the pad assembly from the
ECG-DEFIB port