4.8.3 Usage Review
USAGE REVIEW will display the list of Rescue Events recorded together with their time stamp.
The list of Rescue Events is given in the following table.
Name of Event
Event
Power ON
Indicates the time when the Paramedic CU-ERT is turned ON.
Power OFF
Indicates the time when the Paramedic CU-ERT is turned OFF
Pads ON
Indicates the time when the multifunction rescue pads are
connected to the Paramedic CU-ERT
Pads OFF
Indicates the time when the multifunction rescue pads are
disconnected from the Paramedic CU-ERT
Armed
Indicates the time when the defibrillating capacitor of the
Paramedic CU-ERT is fully charged.
Disarmed
Indicates the time when the charge in the defibrillating
capacitor is dumped on the internal dump resistance of the
Paramedic CU-ERT. The charge in the defibrillating capacitor is
automatically dumped when the SHOCK button is not pressed
within 15 seconds after the prompt to press the SHOCK button
is given.
Shock # X
Indicates the time when the X’th shock is delivered by the
Paramedic CU-ERT.
Analyzing
Indicates the time when the Paramedic CU-ERT is analyzing the
ECG of the patient.
No Shock Advised
Indicates the time when the Paramedic CU-ERT issues the “NO
SHOCK ADVISED” prompt.
Shock Advised
Indicates the time when the Paramedic CU-ERT issues the
“SHOCK ADVISED” prompt.
Paused for CPR
Indicates the time when the Paramedic CU-ERT pauses to give
way to CPR administration
ECG Pads On
Indicates the time when the ECG monitoring cable and
connector assembly is connected to the Paramedic CU-ERT
ECG Pads Off
Indicates the time when the ECG monitoring cable and
connector assembly is disconnected from the Paramedic CU-ERT